Voltage Surge Suppressor into a grounded electrical outlet. Flip the switch on the
Transient Voltage Surge Suppressor to ON position.
Installation is now complete.
Important:
after the unit has been operating for a few hours, unplug the unit and
check all hose connections for leaks (the top hose in particular).
Repeat this
procedure periodically.
Proceed to system operation for further instruction to ensure optimal
performance.
Flushing Instructions
Flushing the system is required after installation or after any disassembly and
cleaning. Flushing may also be required to remove colored water from the unit.
Most filters (if installed) also require flushing prior to use – follow the
manufacturer’s recommendations.
The system may be flushed in two ways. It can be done manually by
disassembling the unit and filling and draining the unit by hand (see Cleaning the
Unit). Flushing may also be done while the unit is operating. Plug in the unit and
open a faucet closest to the unit and run the water for a minimum of 15 minutes.
Control Interfaces
Remote Wireless Monitor (optional)
The Hallett system is available with a remote monitor that monitors the system
operation and alerts you to any change in status. If the unit did not come with a
Remote Monitor, the main unit will display “Radio Initialize Problem” or “Radio
Device Not Installed” upon start up – please ignore this message. If the unit is
equipped with a Remote monitor, install the 9v battery (included) in the remote
monitor and place the monitor in a location of your choice. The remote will
operate up to 164 feet (50m) from the Hallett system location, (Please note that
masonry walls decrease the distance the remote can communicate with the
Hallett system). The remote will run through a diagnostic and then sync with
your Hallett system and display a message indicating unit status. If a warning or
alarm is displayed, detailed information will be available at main unit. If the
remote loses communication with the Hallett, it will display “No Sync”. Move the
monitor to a different location until communication has been reestablished.
The radio frequency used by the remote is 2.4GHz and has FCC modular
approval grant to meet FCC Part15, EN300 328-1, EN301 489-1, Industry
Canada RSS-211.
Setup procedure
Step 1:
Using the Setup Menu on the main UV Pure unit, select “Bind Remote
Device” and choose a bind number (This is a number from 0 to 40 to differentiate