background image

Hallett 15xs Instruction Manual 

Released on 11-Jul-13 – H800003Q

 

Page 29 

UV Pure Technologies Inc. © 2013

 

 

A

LARMS 

&

 

T

ROUBLESHOOTING 

M

ENU 

(

CONT

D

)

 

 

Alarms and 

Troubleshooting

Treatment  Alarm

Review both UVI                

& UVT values

Low UVT Value

Check Pretreat-   

ment devices

Inspect unit              

for fouling

Confirm Wiper    

motor operation

Confirm purge 

valve operation

Check reflectors

Sensor may                

have drifted

New sensor or 

calibration reqd

Lamps require 

replacement

Install 2 lamps                  

& reset counter

Front Panel Ajar

Fully Close               

Front Panel

Wiper not                         

turning

Check wiper               

motor operation

Water                     

temperature high

Check purge              

valve operation

Check shutoff          

valve operation

Circuit board        

temperature high

Check CB fan 

operation

Check if room            

temp within spec

UV Chamber             

Temperature High  

Check if filter             

dirty, fan fail

Check if room          

temp within spec

Temperature                        

Sensor Failure

Check All                    

Temp.  Values

Microprocessor          

failure

Unplug unit and 

recheck.

UV Sensor                

Failure

Check connection

Power reset

Power interrup-        

tion occured

Test Shutoff                      

valve

Confirm valve         

closes and opens

ALARM

HELPFUL      
TIP

WARNING

WARNING

WARNING

WARNING

ALARM

ALARM

ALARM

WARNING

This column represents all possible 
reasons for the fault.

 

Содержание Hallett 15xs

Страница 1: ...UV Pure Technologies Inc ULTRAVIOLET WATER PURIFICATION SYSTEM INSTRUCTION MANUAL MODEL Hallett 15xs VOLTAGE SERIAL NSF ANSI 55 CLASS A CERTIFIED Certified to IEC PUBLICATION 60335 1 EDITION 4 th...

Страница 2: ...uct meet IEC 60335 1 Edition 4 and CE NOTICE THANK YOU By purchasing the Hallett UV system with Crossfire Technology you can now be certain that microbiologically safe drinking water is provided to ev...

Страница 3: ...t UV sensors mounted in air which are designed to not foul and are more reliable indicators of system performance than ordinary systems The lamps are fan cooled and maintain consistent levels of UV ou...

Страница 4: ...can be placed in a convenient location to provide operational information about the Hallett The Hallett 15xs has a built in purge valve The purge valve eliminates the need to shut down the unit in the...

Страница 5: ...ternal Alarm Contacts 17 Remote Start Stop 17 3 OPERATING INSTRUCTIONS 20 Unit Functions 20 Pushbuttons 20 Starting the Unit 21 Remote Wireless Monitor optional 22 Automatic Quartz Cleaning Device 22...

Страница 6: ...ded for the water to be tested for UV transmittance in any applications using cisterns surface water or ground water under the influence of surface water Total Dissolved solids TDS must be less than o...

Страница 7: ...it if water is present on the unit WARNING Ultraviolet Light Hazard The lamps in the unit emit ultraviolet UV light that can cause permanent damage to the skin eyes Never look at the lamp when it is o...

Страница 8: ...Pipe insulation is to be wrapped around the top hose and any other piping overhead to prevent condensation from dripping back onto the unit Note Teflon tape and thread sealant are not required to conn...

Страница 9: ...r Having a Lamp Alarm is normal with a new system or with newly installed lamps until the lamps have reached full power Summary of Installation Unpack and install Hallett on wall Install top bottom ho...

Страница 10: ...35mm 35 76 908mm TO MOUNT USE 12 OR 1 4 DIA SCREW 4 PLACES WATER OUT 1 MNPT WATER IN 1 MNPT 7 54 191mm 23 75 603mm CLEARANCE REQ D TO REMOVE WIPER 9 0 228mm Hallett N SF AN SI55C L ASS A CERTIF IED CE...

Страница 11: ...not flow backwards through the valve See Figure 1D Step 4 Connect the solenoid valve to the Stainless flexible hose attached to the bottom of the Hallett Do not bend the hose excessively Make sure the...

Страница 12: ...STIC SUCH AS PEX DO NOT USE PVC Figure 1B Step 6 Connect the purge valve to a drain using the tubing provided The tubing can be run down the right hand side of the unit within the recess provided the...

Страница 13: ...OCATION OF S N PURGE VALVE RELOCATED BETWEEN OPTIONAL SOLENOID VALVE AND UNIT PURGE VALVE DISCHARGE LINE OPTIONAL DRAIN VALVE HALLETT UNIT MOUNTED VERTICAL AGAINST A WALL GROUNDED OUTLET 115VAC 50 60H...

Страница 14: ...n from falling onto or into the unit Parallel Installation When more than one unit is installed in parallel flow split between units the units must be installed with manual shutoff valves both Hallett...

Страница 15: ...DRAIN MANUAL SHUTOFF VALVE ESSENTIAL SOLENOID VALVE MANUAL SHUTOFF VALVE ESSENTIAL CHECK VALVE ESSENTIAL Figure 1F Step 2 Plug the female end of the power cord into the power entry module located on...

Страница 16: ...eless Monitor optional The Hallett system is available with a remote monitor that monitors the system operation and alerts you to any change in status If the unit did not come with a Remote Monitor th...

Страница 17: ...t the battery on the remote monitor and then reconnect It will then sync up with main unit within a few moments External Alarm Contacts The Hallett system provides an external dry contact for remote a...

Страница 18: ...not restrict movement Step 7 Replace blower and blower housing and allow connector wire back to the circuit board Secure with screw Ensure wires do not obstruct blower Step 8 Plug in blower connector...

Страница 19: ...logies Inc 2013 VIEW A BLOWER HOUSING PANEL FLIPS DOWNWARD SCREW LOOSEN SCREW TO OPEN BALLAST ENCLOSURE DOOR DO NOT REMOVE CIRCUIT BOARD KNOCKOUT LOCATED HERE TILT FORWARD THEN REMOVE AIR FILTER Figur...

Страница 20: ...display four pushbuttons and three indicating lights on the front panel See Figure 2 The pushbuttons allow you to navigate the menu screens on the LCD display See Menu section LCD Display The 2 line b...

Страница 21: ...Hallett contains a safety switch that will disable the lamp power supply ballast if the front panel is not completely closed Caution Avoid continuously starting and stopping the unit within a 24 hr pe...

Страница 22: ...information about the fault would be available at main unit The remote device is battery powered 9V and can be conveniently located near the kitchen tap for a home or in the office for a school Press...

Страница 23: ...onal residences or during a vacation If the possibility of freezing exists the unit and any filters must be drained See Draining the Unit Upon returning reconnect all fittings close all valves and tur...

Страница 24: ...ter and let the bleach remain in the lines for at least 12 hours turn off the Hallett unit Step 7 After the waiting period is over plug in the unit Once stable flush every line for at least five minut...

Страница 25: ...g unit or defeating any system sensors may jeopardize the safety of the water If any system failure occurs and water enters the plumbing system without being disinfected if the optional shutoff valve...

Страница 26: ...cm 2 Net UVT UV Int XX XX XX XX Service Menu See Following Sections Setup Menu Hallett 15xs See Following Sections Alarms and Troubleshooting See Following Sections Message History See Following Secti...

Страница 27: ...eg C deg F UV Air Temp deg C deg F CB Air Temp deg C deg F Audio Alarm AUTO Audio Alarm AUTO Disabled 24hrs 48hrs 1 week OFF Technical Service Enter Password LampStarts Total Last 24hr Power Ups Seria...

Страница 28: ...Continue to operate lamps UV Air Temp High Check if filter dirty fan fail Check if room temp within spec UV Air Temp Low Check if room temp within spec Switch lamps examine values Only one lamp may b...

Страница 29: ...ully Close Front Panel Wiper not turning Check wiper motor operation Water temperature high Check purge valve operation Check shutoff valve operation Circuit board temperature high Check CB fan operat...

Страница 30: ...ng through the unit The lamps in your unit will decay over time and should be replaced every 12 months for optimum performance Your unit has an internal timer to keep track of the lifetime of the lamp...

Страница 31: ...ormation is required review it first before proceeding to Step 1 WING NUT OR THUMB SCREW UNPLUG UNIT TO OPEN UV LAMP LAMP SOCKETS USE A SCREW DRIVER TO PRY LAMP UP HERE w U V w V U FACES IN FACES OUT...

Страница 32: ...he unit being careful not to touch the bulb Ensure that the lamps are aligned with the tab facing the outside of the system left or right side of the system or if the lamps have a stop sign printed on...

Страница 33: ...steps below to inspect the quartz and disassemble the unit for quartz cleaning Caution Always shutdown and unplug the unit before performing any maintenance Never operate a unit unless the front pane...

Страница 34: ...ck and easy procedure that works well in most cases Step 1 Fill a bucket or container with water before shutting off the water supply since you will need the water later to clean the quartz sleeve A s...

Страница 35: ...tt 15xs Instruction Manual Released on 11 Jul 13 H800003Q Page 35 UV Pure Technologies Inc 2013 17 17 17 20 8 8 18 16 16 11 17 12 5 7 13 2 3 15 19 17 10 6 4 1 14A 13 9 14B Flow Restrictor Figure 4B Ha...

Страница 36: ...00031 1 Seal Kit complete 18 R400033 1 Lamp Socket Wire Harness 19 R400047 1 Bottom SS Manifold 1 20 E500021 E500021U 1 Solenoid shut off Valve for 120V systems 240V systems Note that coil is 24Vac in...

Страница 37: ...y and check for leaks Step 11 Replace all covers and plug in the Hallett unit Disassembling the Unit NOTE This procedure is not recommended for individual household users please contact your certified...

Страница 38: ...e bottom quartz seal assembly does not require removal to clean the quartz so leave it in place This will support the quartz sleeve during cleaning and simplify the overall process Step 5 Remove the t...

Страница 39: ...move the quartz sleeve by removing the top and bottom O rings Lift the quartz sleeve out of the unit Step 7 Install the new quartz sleeve into the unit and center it vertically Be careful not to chip...

Страница 40: ...ALIGN HOLE IN SHAFT ADAPTER WITH THE WIPER SHAFT Replacement Parts Use only genuine UV Pure Technologies parts when servicing your Hallett disinfection system Failure to use genuine UV Pure Technologi...

Страница 41: ...nd parts such as replacement UV lamps solenoid valves and other components sold or certified by UV Pure Technologies Inc with which this warranty is included in or on the package What the period of co...

Страница 42: ...rts at its sole discretion and cost with the exception of shipping and handling charges Replacement parts or systems may be functionally equivalent reconditioned refurbished pre owned or new products...

Страница 43: ...r poorly or improperly tightened plumbing connections This Limited Warranty is void if the Product or Parts have been altered or modified in any way by anyone other than a UV Pure technician or a Cert...

Страница 44: ...is the global gold standard accepted by environmental regulatory agents worldwide for water treatment and disinfection This Class A system conforms to NSF ANSI 55 for the disinfection of microbiologic...

Отзывы: