US-4011
.
32
.
8.6.
Communication Details when
USER-COM1
/
2-LAYOUT
is set to
SINGLE
:
8.6.1.
Commands and response
8.6.1.1.
Command:
W<CR>
(57h 0dh), request current reading
Response:
①
<LF>^^^^^^^^U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---over capacity
②
<LF>_ _ _ _ _ _ _ _ U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---under capacity
③
<LF> - - - - - - - - U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---zero-point error
Note:
U
1
U
2
U
3
U
4
U
5
is 1,2,3 or 5 bytes according to current unit: %, kg, lb, pcs, lb:oz
④
<LF><P>W
1
W
2
W
3
W
4
W
5
<DP>W
6
U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---normal data
Note:
(1) The decimal point position is determined by
CONFIG-PRIM.D
(2) If current unit is “lb:oz”, the format will be similar with following:
<LF><P>W
1
W
2
W
3
lb<SP>
W
4
W
5
<DP>W
6
oz
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>
8.6.1.2.
Command:
S<CR>
(53h 0dh) , request current status
Response: <LF> H
1
H
2
H
3
H
4
<CR><ETX>
8.6.1.3.
Command:
Z<CR>
(5ah 0dh)
Response: Zero function is activated (simulate
ZERO/ON/OFF
key) and it returns to current scale status.
<LF> H
1
H
2
H
3
H
4
<CR><ETX>
If ZERO function cannot be activated, it will return to current scale status.
8.6.1.4.
Command:
T<CR>
(54h 0dh)
Response: TARE function is activated (simulate
TARE/PRESET
key), and then returns scale status.
<LF> H
1
H
2
H
3
H
4
<CR><ETX>
If TARE function cannot be activated, it will return to current scale status.
8.6.1.5.
Command:
U<CR>
(55h 0dh)
Response: Changes units of measure (simulate
UNIT/DATA
key) and return scale status with
new
units.
The new measure unit should be allowed to use
<LF> U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>
8.6.1.6.
Command:
L<CR>
(4ch 0dh)
Response: If Hold function can be activated, it will enable/disable hold function (simulate
HOLD/SETUP
key),
and returns scale status.
<LF> H
1
H
2
H
3
H
4
<CR><ETX>
8.6.1.7.
Command:
X<CR>
(58h 0dh)
Response: power off the scale, just like press down the
ZERO/ON/OFF
key to turn off the scale.
8.6.1.8.
Command: all others
Response: Unrecognized command
<LF>? <CR><ETX>
8.6.2.
Summary of Command and Response:
Command
Response
ASCII
HEX
W<CR>
57 0d
Read scale weight:
①
<LF>^^^^^^^^U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---over capacity
②
<LF>________U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---under capacity
③
<LF>-------- U
1
U
2
U
3
U
4
U
5
<CR><LF> H
1
H
2
H
3
H
4
<CR><ETX>---zero-point error
④
<LF><p>W
1
W
2
W
3
W
4
W
5
<dp>W
6
U
1
U
2
U
3
U
4
U
5
<CR><LF>H
1
H
2
H
3
H
4
<CR><ETX>---normal
data
S<CR>
53 0d <LF> H
1
H
2
H
3
H
4
<CR><ETX>; read scale status
Z<CR>
5a 0d <LF> H
1
H
2
H
3
H
4
<CR><ETX> ; simulate ZERO/ON/OFF key