US-4011
.
24
.
and
ZERO/ON/OFF
at same time until first digit flashes, this means indicator has entered into “weight fine-tune”
mode.
5.7.2.
Use
to input correct weight (1230.0). After confirmed by
TARE/PRESET
, the active correct weight
will be displayed and no any digit will be flashed. After this, displayed weight will be adjusted by this ratio
(1230.0/1234.5) and this ratio will keep active until next modification on it.
5.7.3.
To remove effect of this ratio, there’re two ways: “5.6.3.1 way” and “5.6.3.2 way”
5.6.3.1 Do standard calibration, like in “6.CALIBRATION”
5.6.3.2 Move away weight on scale, Press
ZERO/ON/OFF
to make 0 displayed, Put a load onto scale, a number will be
displayed, suppose it’s 1230.0lb ; Press down
TARE/PRESET
and
ZERO/ON/OFF
at same time until first digit
flashes, this means indicator has entered into “weight fine-tune” mode. Press
SETUP
key, the displayed weight
will be restored to 1234.5, and then press
TARE/PRESET
to confirm and exit to normal weighing mode.