Figure 4-III Time Tag Explorer Logic Tab: showing single and coincidence counts
Output Tab
This tab lets user access the Outputs 1-3 (configuration A) or Output 1 (configuration B) on the
device. UseLogic must be checked off in logic tab first in order to access Output functions.
Output Width/ns
Enter the value of desired pulse high time in ns.
Output Event Count
Enter the number of events you expect to occur in one 5 ns time slice under worst-case
conditions. Increasing the value by one will increase the delay of the outputs by 10 ns.
Output Coincidence Pattern
Set coincidence pattern by clicking on the button to change the state of the corresponding
input.
+ The input must have an active edge in the coincidence window for the event to be counted.
- The input must not have an active edge in the coincidence window for the event to be
counted.