w w w . u q d e v i c e s . c o m
L o g i c 1 6 U s e r M a n u a l
| 31
4.2.2
UQD Logic16 Correlation Viewer
Correlation Viewer provides a general interface where you can access Logic Mode directly. This
is helpful if you require coincidence counting.
Figure 4-IX UQD Logic16 Correlation Viewer
Setup of UQD Logic 16 Correlation Viewer
File location: CD Folder>applications>uqd_apps> UQD_correlation_viewer_V0.31_Bin_>
UQD_correlation_counter.exe
Using UQD Correlation Viewer
Parameters
Adjust Input parameters here by changing the following functions.
Tune channel #
Enter which channel you would like to adjust (You also do this by clicking the channel in the
Channel Setting
display).
Threshold [V]
Enter the threshold level in voltage of your input. Click
All
to apply the change to all inputs.
Delay [ns]
Enter amount of time to delay an input signal. Click
All
to apply the change to all inputs.
Polarity
Select which edge of the signal you will like to detect,
Positive Edge
(rising) or
Negative Edge
(falling). Click
All
to apply the change to all inputs.