5.8 Dioden-Test
Prior to any diode test ensure that the diode to be test-
ed is not live. Failure to comply with this prescription
can lead to injuries or instrument damage. Additionally,
foreign voltages falsify the measurement result.
Resistors or semiconductor paths located in parallel
to the diode falsify the measurement result.
"
Select the
measurement range using the meas-
urement function selection switch.
"
Connect the black test lead to the COM socket and
the red test lead to the
socket.
"
Select the diode test pressing the
key. The
message [
diod
] is briefly displayed – connect test
leads to UUT.
"
Read the measurement result displayed.
The multimeters UNITEST Hexagon 710 and Hexagon
720 are equipped with an acoustic socket control fea-
ture (ABK) [ASC]. When wrongly connecting a test
lead, an acoustical signal is audible and the message
InERR
is displayed.
58
Dioden-Test