Transport, setup and acceptance
22
5.2 Ambient conditions
ATTENTION
Ambient conditions during storage and set up.
Observe the ambient conditions! Contact Union Instruments GmbH if the process
gas analyser has been stored for more than three months or needs to be operated
or stored under ambient conditions other than those specified.
5.2.1 Storage conditions
Freezing condensate water in the process gas analyser can cause defects! Protect
the process gas analyser from frost during storage.
Ambient temperature:
-20 - 60°C
Humidity:
0 - 95% relative humidity
Ambient pressure:
700 - 1400 hPa (0.7 - 1.4 bar)
5.3 Installing and connecting
5.4 Setup site
The place of installation of the process gas analyser must meet the following
requirements:
•
Clean dry room (with the exception of INCA5000/INCA6000)
•
No direct exposure to sun
•
Protect from climate influences with a heater or air conditioning if necessary
•
Insure a clean, sufficient amount of ambient air for undistorted measurements
•
Ensure that the load-bearing capacity of the wall is sufficient
WARNING
Leaking process gas can pose a hazard and needs to be discharged by the
operator into a safe environment.
Содержание INCA4004
Страница 1: ...Apr 2016 V1 10 Translation of the original operating instructions Process gas analyser INCA4004 ...
Страница 3: ...3 Dimensions ...
Страница 4: ...4 ...
Страница 8: ...8 ...
Страница 12: ...EU Declaration of Conformity 12 ...
Страница 32: ...Transport setup and acceptance 32 ...
Страница 34: ...Startup switching on 34 ...
Страница 48: ...Operation 48 ...
Страница 50: ...Decommissioning switching off 50 ...
Страница 54: ...Maintenance 54 ...
Страница 58: ...Service 58 ...
Страница 60: ...Accompanying documents 60 ...
Страница 62: ...Disposal 62 ...
Страница 64: ...Spare parts 64 ...
Страница 65: ...65 16 Annex ...
Страница 66: ...Annex 66 EU Declaration of Conformity Flame arrester ...
Страница 67: ...Annex 67 ...