Maintenance
53
Maintenance/Replacing components
Interval
(recommended)
Half-yearly service and after
commissioning
Check and, if necessary, update firmware
version
every 6 months
Save the current configuration with INCACtrl
every 6 months
Annual service
Replace integrated filters
every 12 months
2-yearly service
Replace pump hoses
every 24 months
Replace flame arrester
1)
every 24 months
8-yearly service
Replace integrated pressure reducer
every 8 years
If necessary
Replace gas-delivering pumps
If necessary
Replace sensors, service life depends on
type, message on display „Service [Type]
age“ or „Service [Type] usage“
If necessary
1)
if installed