Operation
58
Main menu - Options - I/O - Digital outputs
Digital signals (relay outputs, floating change-over contacts) are configured in this
menu.
The following must be observed for this:
Zero position:
The zero position of the digital outputs can be freely selected: low/high.
Process:
The process gas solenoid valve is open and the flame is burning.
Calibration:
The calibration gas solenoid valve is open and the flame is burning.
Maintenance:
The device must be checked in a foreseeable time, but is still ready for operation.
The reason for maintenance can be obtained from the "General information" field.
Examples of reasons include:
•
Filter change (air filter)
•
No calibration gas (calibration canceled)
•
Room temperature too high (air inlet temperature > 41 °C)
•
Gas pressure too low
•
Calibration deviation outside of tolerances
•
The calibration was unstable and was canceled
Содержание CWD2005 SPC
Страница 1: ...Apr 2016 V1 06 Original Operating Instructions Combustion Calorimeter CWD2005 CWD2005 Plus ...
Страница 5: ...Technical data 5 1 Technical data 1 1 Dimensions Fig 1 1 Housing dimensions Weight approx 54 kg ...
Страница 10: ...Technical data 10 ...
Страница 12: ...EU Declaration of Conformity 12 ...
Страница 18: ...Safety information 18 ...
Страница 22: ...Protective equipment 22 ...
Страница 32: ...Transport installation and acceptance 32 Fig 6 1 Wall mounting Fig 6 2 Dimensions wall mounting ...
Страница 48: ...Description of the work stations operator control elements 48 ...
Страница 63: ...Operation 63 Main menu Options Calibration Calibration values are configured in this menu ...
Страница 76: ...Operation 76 Main menu Options System Settings Change color The display colors can be changed here ...
Страница 87: ...Operation 87 ...
Страница 88: ...Operation 88 ...
Страница 90: ...Decommissioning Switching off 90 ...
Страница 94: ...Maintenance 94 ...
Страница 100: ...Troubleshooting 100 ...
Страница 102: ...Service 102 ...
Страница 104: ...Disposal 104 ...
Страница 106: ...Spare parts 106 ...