Operation
57
Main menu - Options - I/O - Analog outputs
The mA signals are configured in this menu.
Signal (list field)
Wobbe-index, density, heating value,
Type (list field)
4 – 20 mA
Unit (list field)
kcal/m³, relative
MB from MB to
value fields for numeric input
The following must be observed for this:
Units:
xxx/m
³ and BTU/ft³ correspond to a gas temperature of 0
°C and a barometric pressure of
1013 mbar.
xxx/Sm
³ corresponds to a gas temperature of 15 °C (60 °F) and a barometric pressure of
1013 mbar.
xxx stands for MJ, kcal, or kWh.
BTU/ft³ corresponds to a gas temperature of 15 °C (60 °F) and a barometric pressure of
1013 mbar.
Содержание CWD2005 SPC
Страница 1: ...Apr 2016 V1 06 Original Operating Instructions Combustion Calorimeter CWD2005 CWD2005 Plus ...
Страница 5: ...Technical data 5 1 Technical data 1 1 Dimensions Fig 1 1 Housing dimensions Weight approx 54 kg ...
Страница 10: ...Technical data 10 ...
Страница 12: ...EU Declaration of Conformity 12 ...
Страница 18: ...Safety information 18 ...
Страница 22: ...Protective equipment 22 ...
Страница 32: ...Transport installation and acceptance 32 Fig 6 1 Wall mounting Fig 6 2 Dimensions wall mounting ...
Страница 48: ...Description of the work stations operator control elements 48 ...
Страница 63: ...Operation 63 Main menu Options Calibration Calibration values are configured in this menu ...
Страница 76: ...Operation 76 Main menu Options System Settings Change color The display colors can be changed here ...
Страница 87: ...Operation 87 ...
Страница 88: ...Operation 88 ...
Страница 90: ...Decommissioning Switching off 90 ...
Страница 94: ...Maintenance 94 ...
Страница 100: ...Troubleshooting 100 ...
Страница 102: ...Service 102 ...
Страница 104: ...Disposal 104 ...
Страница 106: ...Spare parts 106 ...