Transport, installation, and acceptance
29
6.3 Installing and connecting
6.3.1 Installation location
The installation location of the combustion calorimeter must meet the following
conditions:
•
Clean room that may be used only for purposes of gas analysis and measure-
ments.
•
At least 50 cm working space on the left side
•
Lockable windows
•
Protected from direct influence by the weather and direct solar radiation
•
Stable room temperature
•
Provided with clean and adequate ambient air for unbiased measurement re-
sult (combustion calorimeter requires approximately 30 m³/h air)
•
Adequate load bearing capacity of the wall is ensured
•
Mounting on a solid wall
Содержание CWD2005 SPC
Страница 1: ...Apr 2016 V1 06 Original Operating Instructions Combustion Calorimeter CWD2005 CWD2005 Plus ...
Страница 5: ...Technical data 5 1 Technical data 1 1 Dimensions Fig 1 1 Housing dimensions Weight approx 54 kg ...
Страница 10: ...Technical data 10 ...
Страница 12: ...EU Declaration of Conformity 12 ...
Страница 18: ...Safety information 18 ...
Страница 22: ...Protective equipment 22 ...
Страница 32: ...Transport installation and acceptance 32 Fig 6 1 Wall mounting Fig 6 2 Dimensions wall mounting ...
Страница 48: ...Description of the work stations operator control elements 48 ...
Страница 63: ...Operation 63 Main menu Options Calibration Calibration values are configured in this menu ...
Страница 76: ...Operation 76 Main menu Options System Settings Change color The display colors can be changed here ...
Страница 87: ...Operation 87 ...
Страница 88: ...Operation 88 ...
Страница 90: ...Decommissioning Switching off 90 ...
Страница 94: ...Maintenance 94 ...
Страница 100: ...Troubleshooting 100 ...
Страница 102: ...Service 102 ...
Страница 104: ...Disposal 104 ...
Страница 106: ...Spare parts 106 ...