Transport, installation, and acceptance
30
6.3.2 Room ventilation
WARNING
Risk of injury due to temperature of exhaust gas/housing!
The exhaust gas temperature is between 8 - 20 °C above the housing tempera-
ture!
Use personal safety equipment to prevent burns!
NOTE
In the case of unfavorable circulation/installation conditions, appropriate deflect-
ing sheets must be provided that prevent fresh air from flowing directly onto the
combustion calorimeter.
The following conditions must be met:
•
No direct fresh air introduction in the combustion calorimeter.
•
Exhaust gases must be discharged to the environment protected from air
draft using a suitable pipe/tube and with the help of corresponding fresh air
mixing. Connection points must be checked for leak tightness.
Содержание CWD2005 SPC
Страница 1: ...Apr 2016 V1 06 Original Operating Instructions Combustion Calorimeter CWD2005 CWD2005 Plus ...
Страница 5: ...Technical data 5 1 Technical data 1 1 Dimensions Fig 1 1 Housing dimensions Weight approx 54 kg ...
Страница 10: ...Technical data 10 ...
Страница 12: ...EU Declaration of Conformity 12 ...
Страница 18: ...Safety information 18 ...
Страница 22: ...Protective equipment 22 ...
Страница 32: ...Transport installation and acceptance 32 Fig 6 1 Wall mounting Fig 6 2 Dimensions wall mounting ...
Страница 48: ...Description of the work stations operator control elements 48 ...
Страница 63: ...Operation 63 Main menu Options Calibration Calibration values are configured in this menu ...
Страница 76: ...Operation 76 Main menu Options System Settings Change color The display colors can be changed here ...
Страница 87: ...Operation 87 ...
Страница 88: ...Operation 88 ...
Страница 90: ...Decommissioning Switching off 90 ...
Страница 94: ...Maintenance 94 ...
Страница 100: ...Troubleshooting 100 ...
Страница 102: ...Service 102 ...
Страница 104: ...Disposal 104 ...
Страница 106: ...Spare parts 106 ...