Operation
61
Main menu - Options - I/O - Digital inputs
Various digital signals are configured in this menu.
The following must be observed for this:
The pin assignment is preassigned.
Chapter 6.3.10 Connector assignment Input-Output IOexternal!
Signals
Zero position
Contact
CWD2005
Start calibration
high
open
Calibration starts
closed
No calibration
low
open
No calibration
closed
Calibration starts
Start measurement
high
open
Measuring starts
closed
Measuring stops
low
open
Measuring stops
closed
Measuring starts
Hold signals
high
open
Hold mA
closed
mA online
low
open
mA online
closed
Hold mA
Cancel calibration
high
open
Calibration cancellation
closed
No calibration cancellation
low
open
No calibration cancellation
closed
Calibration cancellation
Содержание CWD2005 SPC
Страница 1: ...Apr 2016 V1 06 Original Operating Instructions Combustion Calorimeter CWD2005 CWD2005 Plus ...
Страница 5: ...Technical data 5 1 Technical data 1 1 Dimensions Fig 1 1 Housing dimensions Weight approx 54 kg ...
Страница 10: ...Technical data 10 ...
Страница 12: ...EU Declaration of Conformity 12 ...
Страница 18: ...Safety information 18 ...
Страница 22: ...Protective equipment 22 ...
Страница 32: ...Transport installation and acceptance 32 Fig 6 1 Wall mounting Fig 6 2 Dimensions wall mounting ...
Страница 48: ...Description of the work stations operator control elements 48 ...
Страница 63: ...Operation 63 Main menu Options Calibration Calibration values are configured in this menu ...
Страница 76: ...Operation 76 Main menu Options System Settings Change color The display colors can be changed here ...
Страница 87: ...Operation 87 ...
Страница 88: ...Operation 88 ...
Страница 90: ...Decommissioning Switching off 90 ...
Страница 94: ...Maintenance 94 ...
Страница 100: ...Troubleshooting 100 ...
Страница 102: ...Service 102 ...
Страница 104: ...Disposal 104 ...
Страница 106: ...Spare parts 106 ...