Manual – Starflow Ultrasonic Doppler Instrument
Model 6526
Unidata Manual - 6526 Starflow User Manual Issue 17 01 2020.docx
Page 74
15.3.3
Measurement Errors
If during a data collection sequence (SDIGET) an error occurs, the sequence is retried a total of
ten times (by default). A non-recoverable error causes the SDIGET instruction to report a “null”
result and a subsequent SDICONV will not return any data (the channel readings remain
unchanged from their last good reading).
15.3.4
Starflow SDI-12 Command Implementation
The following table shows which SDI-12 commands are and are not recognised:
Command
Description
Sensor
Data
Recorder
a!
Acknowledge
X
aI!
Send identification
X
aV!
Start verification
X
aM!
Start measurement
aD0!
Send data
S
aM1...9!
Start additional measurement
X
aD1...9!
Send additional data
X
X
extended commands
X
15.4
SDI-12 Test
The
SDI-12 Test
function allows you to check the operation of the data logger as
an SDI-12 recorder and the operation of any SDI-12 sensors connected to it. You
are able to issue identify, measurement, and read commands.
When the window is first opened, click
Connect
to connect to the data logger and
start SDI-12 diagnostics: