6.1.2.4 Configuring the Starflow Advanced options
The advanced option TAB contains very important factors which are used to interpret the Starflow
data. The table below shows the default settings set by the software for general flow operation:
Options likely to be configured:
Min operating Depth:
The Starflow minimum operating depth is set to 25mm this is the minimum
depth which is required in order for the Starflow to operate in the correct manner.
Max Quartile Setting
: The quartile setting is used to determine how accurate the data being
reported is and if the Starflow is picking up good signals or bad signals, if a quartile higher than
100 is generated this generally says that the data being recorded is unreliable.
Min Samples to analyse:
The Starflow requires a certain number of samples to determine if the
data being provided is accurate, the default setting is 20 and this is recommended, if there is less
than this then the data could be widely spread causing inaccuracies in reported data.
Max number of samples:
This setting allows the Starflow to receive a certain number of samples
during a scan rate period, in general the more samples the better as then your range of velocity
samples becomes more accurate, in general circumstance 200 samples is enough for good data
quality but in poor flow areas this should be increased to 500.
Set to zero or use last value: If one of the CDT criteria is not met (e.g. quartile > 100 or Min
samples to analyse < 20 ) then the software will either record a zero or the last value obtained, in
general flow conditions and for data quality a zero is usually advised.
The other factors are factory defaults and should not be changed.