Uni-Trend Technology (Chengdu) Ltd.
Page 24 of 27
Communication & Control Software for UTD2025_3025 Series Oscilloscopes User Manual
3.3
GPIB interface controls
The GPIB port is a dedicated interface module of UTD3025 Series Digital Storage Oscilloscopes. After installing
the communication and control software, you can carry out system operation and control your oscilloscope
through the GPIB interface. You can create an automatic testing system simply, quickly and efficiently, and carry
out testing and measuring.
With the optional UT-M02 extension module, you can operate your UTD3025 oscilloscope through the GPIB
interface as follows :
1.
Select “GPIB” in the equipment selection option on the real-time control panel.
2.
When “GPIB” is selected, the dialog box in Fig 3-6 will pop up.
Fig 3-6
Enter the GPIB device address in this box (range : 1-30). The GPIB device addresses set in the software and the
communicated oscilloscope must be the same, otherwise communication between the oscilloscope and the
software cannot be established.
Access sub-menu page 3 by pressing [UTILITY]
GPIB on the oscilloscope control panel for GPIB device
address setup.
There must not be two identical GPIB device addresses on one GPIB main line.
3.
For other operation instructions refer to steps 2 and 3 for “USB interface controls” in this chapter.
Note :
When using the GPIB interface, the “transmission interface” feature is disabled.