ULVAC CRYOGENICS INCORPORATED
11
⑫
Compressor alarm light
[Conditions that alarm light illuminates]
(1) When overcurrent by compressor overload is detected
(2) When compressor wiring is interrupted
⑬
Run/Stop button and light
Press this button to startup or stop the equipment. The button illuminates while
running.
The light flashes when the compressor has been running for 8000 hours (Inspection
interval).
⑭
Reset button
Resets equipments when the compressor trouble or abnormal pressure occur.
⑮
Pressure regulator (inside)
Adjusts the discharge pressure of produced gas. Set the value within specified
range referring to the nitrogen discharge pressure gauge.
⑯
Pressure sensor (inside)
Detects the tank pressure. This sensor triggers equipment emergency stop when the
pressure is below the setting while nitrogen is being discharged.
⑰
Circuit breaker
Circuit breaker for the entire equipment.