SK00-9663-EI-003-00
25
5.3.
I/O input signal (Analog output type SW100-A only)
From the I/O connector of this unit, perform zero point adjustment input and atmospheric
pressure adjustment input.
Internally, it is connected to the power supply voltage (DC 14V to 30V) input to this
unit, so use a contact capacity of DC 30V or higher, or a power supply voltage or higher.
When using it, make a short cut between the pin of the signal to be operated and the
GND terminal. For details on the adjustment method, refer to Chapter 8 of this manual.
*
CAUTION
*
Contact capacity
The capacity of the externally installed contacts should be equal
to or higher than the input power supply voltage or DC 30V or
higher.
*
CAUTION
*
Contact leak current
Pay attention to the leak current of the contacts. If a current
of 0.1mA or more flows between the input signal pin and the GND
terminal, it may be treated as if the signal was input.
Diagram of input signal internal circuit
Power supply
+5 V
5:ADJUST input
15: GND
Internal circuit