Safety instructions 

 

Electrical 

Connect to a 220-240 VAC fully earthed supply via a 3pin plug. 

It can be dangerous to operate an ultrasonic bath without an earth connected. 

The mains plug is fitted with a 5A fuse. 

NEVER FIT A FUSE OF A HIGHER RATING. 

Ensure that excess mains cable is stored neatly. 

General use 

Ensure that the bath contains liquid before you switch it on. 

Always  fill  with  cold  water.  This  will  ensure  that  the 

Hygea  6427

  always  runs  its 

automatic de-gas cycle (see pg 29 for the importance of degassing). 

Do not place fingers or other parts of the body in the bath. 

Always use a basket to support items to be cleaned.  

Not doing so may damage the bath and invalidate your warranty. 
 

Care  should  be  taken  when  operating  the  bath  at  higher  temperatures  as  external 
surfaces may become hot.  

For  best  cleaning  results  use  a  detergent,  at  a  suitable  dose,  recommended  for  use  in 
ultrasonic  cleaning  baths.  Never  use  toxic,  flammable,  acidic,  caustic  or  corrosive 
solutions in the bath. 

Refer  to  the 

Hygea  6427

 

Operator  Instructions 

manual  for  full  guidance  on  the  correct 

use of the bath. 

The  user  should  familiarise  themselves  with  this 

Operator  Instructions 

manual  before 

operating the equipment and should apply to Ultrawave or its agent for advice on cleaning 
techniques  or  detergents.  Ultrawave  will  not  be  responsible  for  damage  or  injury  caused 
by incorrect use of the equipment. 

Ultrawave  will  not  be  responsible  for  damage  or  injury  caused  by  incorrect  use  of  the 
equipment. 

 

Содержание Hygea 6427

Страница 1: ...Hygea 6427 Operator Instruction Manual...

Страница 2: ......

Страница 3: ......

Страница 4: ......

Страница 5: ...intenance Cleaning 21 Periodic testing and validation 22 Validation Testing Reminder 23 Hygea 6427 Cycle validation 24 Using the Ultrawave Cycle Validation Software 25 Hygea 6427 good practice 30 Dete...

Страница 6: ...so may damage the bath and invalidate your warranty Care should be taken when operating the bath at higher temperatures as external surfaces may become hot For best cleaning results use a detergent at...

Страница 7: ...card 1 x Hygea 6427 Administrator instruction manual 1 x Hygea 6427 Operator instruction manual How to install The Hygea 6427 should be installed on a benchtop or stand if provided located close to a...

Страница 8: ...ated to the minimum operating temperature the Hygea 6427 is ready to use although an extra degas cycle may be required Select your operator identification Press the ENTER key to start the cleaning cyc...

Страница 9: ...stainless steel clips To use these press the small stainless clip in towards the pipe and pull the connector straight up the connector is then pushed into the other receptacle 1 The tubes can be conne...

Страница 10: ...tor into the tube and then connecting the hollow instrument 5 The instrument and hose can then be connected to the manifold Important The tubes must be connected whenever the Hygea 6427 is in use not...

Страница 11: ...first switched on the following screen will appear for approximately 5 seconds While this screen is displayed the buttons are inoperative This is normal as the Hygea 6427 is running through its start...

Страница 12: ...wing screen will display The Hygea 6427 will automatically start heating and degassing the liquid When the liquid reaches the required temperature the following screen will be displayed and the Hygea...

Страница 13: ...down the list indicating which name is selected Press the Next button to show the next group of four names Up to 20 names can be stored in the operator list When the correct operator is selected pres...

Страница 14: ...ays to confirm that detergent has been added Press the Start button to confirm that the liquid contains detergent See page 30 for information about detergents The Hygea 6427 will then begin the cleani...

Страница 15: ...d 11th minutes If the pump time is set to 2 minutes the pump will run during the 1st and 2nd 6th and 7th and 11th and 12th minutes While the pump is in operation the word pump will display on screen W...

Страница 16: ...ng the details of that cycle Once the ticket is printed simply tear off and store as necessary Ultrawave Version V3 00 Model HYGEA 6427 Serial No 056204 Time 15 30 Date 01 12 17 Hops A Op 01 Nurse A C...

Страница 17: ...on page 12 to the MC card This information can then be transferred to a PC and an electronic copy saved using the Ultrawave Cycle Validation software supplied on the MC Card It is recommended that th...

Страница 18: ...e screen displays a Low Fluid warning as below the bath will not operate until the unit is filled to the correct level and the detergent dose added If the screen shows a Lid Open warning the cycle wil...

Страница 19: ...ng press and hold the ENTER button at the bottom of the screen until the warning stops If the operator aborts the cleaning cycle the following screen will be displayed If the fluid level drops below t...

Страница 20: ...d on the MC card when the machine is next switched back on The cycle will be recorded as an aborted cycle If the cycle fails due to a power failure the following screen is displayed when the power is...

Страница 21: ...lowing screen will display If the ultrasonic power drops below a certain point during a cycle the cycle will stop and the following screen will be displayed If the cycle aborts the reason will be disp...

Страница 22: ...sure that the drain is either directly over a sink or is connected to a tube which is placed into a sink Step 2 Turn the tap until open DRAIN TAP SHOWN OPEN Step 3 Allow the liquid to drain Step 4 Clo...

Страница 23: ...new paper rolls need to be ordered Use the paper feed button to feed through the last few centimetres of paper and fit a new roll Step 1 Open the printer paper tray by pulling the clear cover towards...

Страница 24: ...g Step 5 Close the cover Step 6 Close the printer paper tray Step 7 Feed a small amount of paper through the printer by pressing the paper advance button on the right hand side of the printer or the p...

Страница 25: ...nfected daily Cleaning and disinfection processes may be regulated by the CSSD HSDU manager or AP so check with these before attempting manual disinfection Check hoses for any signs of visible blockag...

Страница 26: ...available from Ultrawave If there is no access to the Activity Meter this test should be replaced by the aluminium foil ablation test 3 Quarterly tests carried out by the operator Estates department U...

Страница 27: ...Therefore after 11 months in operation a reminder will appear on the LCD display warning that the annual validation testing is due The Validation Testing reminder will display as below When this remin...

Страница 28: ...more information If the hard copy print out is chosen then at the end of each cycle the integral printer in the Hygea 6427 will print a ticket showing the details of that cycle Once the ticket is prin...

Страница 29: ...rd the card must be inserted into the port located on the left hand side of the control panel before your Hygea 6427 is switched on The digital cycle traceability will save the information of each cyc...

Страница 30: ...IT administrator Also ensure that Microsoft Net 3 5 Framework is installed as well as the associated updates In order to install the Software the MC Card should be inserted into the PC using the USB c...

Страница 31: ...lready saved to the PC the data will display in the below format If you wish to copy new cycle data to the software click on the Copy Cycle Results from MC Card button You will need to ensure that the...

Страница 32: ...by pressing the print button It is also possible to print data from multiple cycles This can be done by clicking on multiple cycles in the left hand column In order to print the data from multiple cyc...

Страница 33: ...in the applicable search box located in the top right of the screen The following fields can be used to search for cycle data Serial Number of the machine Location Clinic Practice Operator name Date...

Страница 34: ...ning medical instruments the maximum temperature should be set to 40 C Setting the Max above 40 C may cause proteins to coagulate and bake on to the instruments Ultrasonic activity itself will heat up...

Страница 35: ...keep your bath clean Not only will contaminated liquid reduce the performance of the bath it may also damage it Change the cleaning liquid regularly Your cleaning process will determine how often to...

Страница 36: ...g is now 80 decibels There is also an exposure limit value of 87 decibels taking account of any reduction in exposure provided by hearing protection above which workers must not be exposed To help you...

Страница 37: ...EN 13202 2000 Ergonomics of the thermal environment Temperatures of touchable hot surfaces Guidance for establishing surface temperature limit values in production standards with the aid of EN 563 199...

Страница 38: ...able acidic caustic or corrosive chemicals or fluids not recommended by Ultrawave The user should familiarise themselves with this instruction booklet before operating the equipment and should apply t...

Страница 39: ...s message displays run another cycle to see if the problem persists The screen displays OVER TEMP Drain and refill with fresh cold water The Hygea 6427 is fitted with a temperature sensor that automat...

Страница 40: ...tect the health and safety of Ultrawave employees reducing the risk of potential injury or infection More information on decontamination guidelines as well as copies of decontamination certificates ca...

Страница 41: ...tor test strips and holders Brownes strips and holders for weekly testing of the Hygea 6427 Service Contracts and Validation Testing Ultrawave can offer full Service and Validation packages for the th...

Страница 42: ..._______________ __________________________________________________________ __________________________________________________________ __________________________________________________________ _______...

Страница 43: ...39...

Страница 44: ......

Страница 45: ......

Страница 46: ...ce Tel 44 0 845 330 4238 Fax 44 0 845 330 4231 www ultrawave co uk admin ultrawave co uk service ultrawave co uk Ultrawave baths are manufactured in the United Kingdom Ultrawave baths are manufactured...

Отзывы: