11
Starting a cycle
The
Hygea 6427
is factory pre-set with a total cycle time of 15 minutes. This consists of
3 consecutive cycle stages of 5 minutes and the countdown will display from 5 minutes.
During this cycle, the irrigation pump will operate three times, once at the beginning of
each cycle stage.
The default pump cycle time is set to 1 minute. Under these parameters, the pump will
run during the 1
st
, 6
th
and 11
th
minutes. If the pump time is set to 2 minutes, the pump
will run during the 1
st
and 2
nd
, 6
th
and 7
th
and 11
th
and 12
th
minutes.
While the pump is in operation, the word “pump” will display on screen:
When the pump is not in operation, the following screen will display: The default
ultrasonic cycle time of the
Hygea 6427
is 15 min. The following screen will appear.
The cycle can be aborted at any time by the operator, by pressing the “Abort Cycle”
button and the cycle time can be changed in Administrator mode. Please refer to pg 11 of
the
Administrator Instruction
manual for information on setting the cycle time.