NT228B GB2
71 / 98
15.1
Saving and recalling of a calibration file:
The UF 801-P has memory space to store 1+2 calibrations / settings.
The current configuration called “Config. 0”
Two (or more) configuration spaces (1 & 2) that you are free to use.
All these operations are done from Configuration menu.
LANGUAGE : ENGLISH
-- STATION’S NAME --
xxxxxxx
---- BACK LIGHT ----
xxxxxxx
---PC CONNECTION ---
N° JBUS : xxxxxx
Bits Rate: xxxxxx
--- OPTION FILES ---
Load Option File
N°*:xxxxxxx
Save Option File
N°*:xxxxxxx
To keep a record of actual settings
(now in config. 0):
Give it a name at Station Name : Site
/
line Nr. (letter by letter)
At Save Option File, select memory space Nr. 1 or 2.
Validate the saved record by changing the line (V).
To recall these archived settings:
Go to “Load Option File” line and select which one you want
To load the file, press the down arrow (change of line)
The corresponding data overwrite those previously in Config. 0, now lost
If you want
to cancel a record
, just overwrite on it with a new one.
These records can be done with Simple / Normal or Expert modes. If you want to
import a normal or an expert setting file, you need to configure your UF 801-P in
the corresponding mode.
Settings Reset:
UF 801-P has a permanent memory of the typical factory
settings. They correspond to our tests, and UF 801-P is normally delivered with
these data as actual settings. To re-install this factory settings file, select
“Settings Reset” and “Yes”
in configuration menu.
Содержание UF 801-P
Страница 40: ...NT228B GB2 40 98 8 DUAL PIPE VERSION UF 801 PB 8 1 Introduction to UF 801 PB 8 2 Complementary screens Contents...
Страница 59: ...NT228B GB2 59 98 12 TOTALIZERS 12 1 Description Contents...
Страница 67: ...NT228B GB2 67 98 14 TIMER SEQUENCER 14 1 Settings Contents...
Страница 87: ...NT228B GB2 87 98 19 BATTERY CHARGER 19 1 Battery Charger 19 2 Battery pack whole life Contents...