NT228B GB2
32 / 98
6.3
Pipe/Fluid settings for the application
Mode Simple considers water in usual ambient temperature range. Sound velocity,
which drives the requested distance between probes, and also signal tracking
window, is considered around
15°C
, with
Co=1472 m/s
with a whole
Delta C
range sufficiently wide at
400 m/s
(+/- 200 m/s).
PIPE/FLUID
SETTING
------- PIPE -------
-------Fluid--------
Product = WATER
--------Tube--------
Ext.Diam. = ******mm
Ext.Circ. = ******mm
--------Wall--------
Material = xxxxxxxx
Thick. = ******mm
Sp.Sound = ****m/s
Please enter with accuracy the pipe dimensions as Ext. Diam or Ext. Circ.(=
*O.D). UF 801-P displays both and calculates the pipe I.D.
Use a tape measure or a vernier for Ext. Diam. and a thickness gage for pipe
thickness unless drawings are available (or marks on pipe / site construction
records).
6.4
Probes choice:
It is very important to
select the right probes & supports
for your application.
> See §5.1. Depending on these probes and your application or pipe conditions,
select the best compromise for the probe installation. > See § 3.4 and 5.2
By
default, start with the V Reflex mode
.
------- PIPE -------
-------Chord1-------
Probe = xxxxxx
Mounting = xxxxxx
------- PIPE -------
-------Chord1-------
Probe (SE) = xxxxxx
Mounting = xxxxxx
Probe (SM) = xxxx
Length = xxxx.x mm
Axial D = xxxx.x mm
(Welter probes menu)
(Clamp-on probes menu)
Содержание UF 801-P
Страница 40: ...NT228B GB2 40 98 8 DUAL PIPE VERSION UF 801 PB 8 1 Introduction to UF 801 PB 8 2 Complementary screens Contents...
Страница 59: ...NT228B GB2 59 98 12 TOTALIZERS 12 1 Description Contents...
Страница 67: ...NT228B GB2 67 98 14 TIMER SEQUENCER 14 1 Settings Contents...
Страница 87: ...NT228B GB2 87 98 19 BATTERY CHARGER 19 1 Battery Charger 19 2 Battery pack whole life Contents...