8.1
Introduction to UF 801-PB
The UF 801 Hardware permits two ultrasonic chords by multiplexing internal
resources of the emitting/receiving circuits. The two chords option can be
designed for installation on a same pipe (UF 801-PB) or on two chords on a same
pipe (double set of clamp-on probes or two parallel intrusive probes.
------ TOTAL Q------
------General-------
Number of pipes= 2
QT Unit = m
3
/h
QT = +/-QA+/-QB
-------Graph-------
QT Period = 5 s
QT Q min = xx.x
QT Q max = xx.x
An UF 801-PB will also work on one pipe with Nr of pipes=1. It will be very similar
to UF 801-P but QT display (as total flow) that remains and will be equal to QA.
After setting
Nr. of pipes =2
(Pipe/Fluid Menu), you will have complete and
totally independent resources to set your applications on two pipes.
The operation QT from QA and QB can be set with coefficients = +1 / -1 or Zero to
do a copy, a sum or a difference.
8.2
Complementary displays :
They are very similar to the previous ones, but mention to A or B :
------ PIPE A-------
-------Chord1-------
Probe (SE) = xxxxxx
Mounting = xxxxxx
Ext.Coeff =Auto/Man
Delta To = x.xx ns
Gain Type =Auto/Man
Gain Max = xxx dB
or Gain = xxx dB
------- PIPE B------
--------Tube--------
Ext.Diam. = ******mm
Ext.Circ. = ******mm
--------Wall--------
Material 1= xxxxxxx
Thick.1 = ***.**mm
Sp.Sound = ****m/s
(
If Other
)
Содержание UF 801-P
Страница 40: ...NT228B GB2 40 98 8 DUAL PIPE VERSION UF 801 PB 8 1 Introduction to UF 801 PB 8 2 Complementary screens Contents...
Страница 59: ...NT228B GB2 59 98 12 TOTALIZERS 12 1 Description Contents...
Страница 67: ...NT228B GB2 67 98 14 TIMER SEQUENCER 14 1 Settings Contents...
Страница 87: ...NT228B GB2 87 98 19 BATTERY CHARGER 19 1 Battery Charger 19 2 Battery pack whole life Contents...