NT228B GB2
44 / 98
9.1
When and How Use this menu ?
ECHO
DISPLAY
This menu function permits you to
check your ultrasonic measurement
conditions and know if they are
good,
acceptable,
weak
or
impossible.
The parameter
Q. max=
permits to adjust the Delta T display sensitivity
depending your application. The display does not show the exact Delta T: It is
limited to ¼ of the ultrasonic signal period. From Q. max setting, you can force
more or less a Delta T display for any small or large actual Delta T.
There are two graphic displays
(4 if a dual chord / pipe version):
The first one shows, from the left to
the right side, all energised signals
(ultrasonic, but also noises) that UF
801-P is receiving and must filter and
treat.
The two dotted vertical lines are
showing the selection window that
result from actual settings (Diam. Or
L and also C)
Any signal out from this window will
be ignored. The beginning must be
clean.
The second screen is an expanded
view inside above window that results
from the digital acquisition in a
tracking window, our electronic lens
for ultrasonics.
This screen shows two dotted lines
when all is OK.
- The vertical one shows triggered
transit time measurement, usually at
zero crossing.
- The horizontal one shows the
automatic threshold result with the
selected edge on the signal.
Содержание UF 801-P
Страница 40: ...NT228B GB2 40 98 8 DUAL PIPE VERSION UF 801 PB 8 1 Introduction to UF 801 PB 8 2 Complementary screens Contents...
Страница 59: ...NT228B GB2 59 98 12 TOTALIZERS 12 1 Description Contents...
Страница 67: ...NT228B GB2 67 98 14 TIMER SEQUENCER 14 1 Settings Contents...
Страница 87: ...NT228B GB2 87 98 19 BATTERY CHARGER 19 1 Battery Charger 19 2 Battery pack whole life Contents...