Page 18
38500 Instruction Manual (Rev. 1)
5.4
Applied force ratemeter
If the
Ratemeter
option is selected in the configure menu, then, during each measurement,
PAM’s display is similar to:
The white rectangle on the second line moves left or right of the centre marker when the
applied force is respectively below or above that required to maintain the selected target
force rate. In the above example, the applied force is below the ideal level for the specified
target force rate.
The fourth line of the display is a time progress bar indicating how much of the available
measurement time has been used.
5.5
Available Configuration Settings
The complete list of configuration options is shown below including the adjustment ranges
and default values.
Configuration flags which may be set or cleared.
Whether each measurement is ended
automatically on the force level
(default) or
by pressing the foot pedal.
Whether
every measurement is stored regardless
(sometimes required by GLP) or
the user may choose which measurements to store
(default).
Whether
the transducer offset value is shown before each measurement
or
not
(default).
Whether
a warning noise sounds if a preset force is exceeded during a
measurement
or
not
(default).
Whether
abbreviated configuration settings are shown on the Measure menu
or
not
(default).
Whether
the stored data includes the force waveform
(default) or
not
.
Whether
the operator should be asked to indicate Left or Right limb
or
not
(default).
Whether
PAM should use Subject Numbers
or
not
(default).
Whether
PAM should use the applied force ratemeter during tests
or
display
live force and time in text
(default).