UGO BASILE - I
TALY
Page 8
38500 Instruction Manual (Rev. 1)
Press
Measure
. A screen will appear similar to:
After a few seconds, the instrument completes an
Autozero
operation and then the display
shows.
IMPORTANT NOTE:
do not apply force while the instrument is autozeroing.
Note that, in this case, all three
soft buttons
are in use.
Press
Select this
. The display is similar to:
Check that no force is being applied to the connected transducer and then press
Test.
PAM checks the transducer zero and then displays:
Apply
gentle
1
force to the transducer. The display changes to typically.
For as long as the force is applied steadily, the
Force
value will indicate the applied force in
gf and the
Time remaining
will indicate the number of seconds left to complete the test.
Release the transducer after a few seconds and the screen changes to something like:
1
Do not apply more than 500 or 1500 gf (mouse or rat transducers respectively). Excessive force can
permanently damage the transducer.
Measure Peak Force
Apply the force NOW
PAM Transducer
Large: 0->1500gf
Calibrated: 091009
PAM gain matched
Measure Peak Force
--------------------
Next View Select
option config this
23/07/09 14:23:35
Test 1
Exit Test