Series 58000 Instruction Manual (Rev.1)
4.3.2
Fuse Holder & Voltage Selector
The fuse holder comprises two fuses, one on the live, and the other on the neutral. For operation
at 230-240 Volts, we recommend 315 mA timed “slow blow” fuses (type T315). Use 630 mA
fuses (type T630) for operation at 115 Volts.
The fuse holder also embodies the Voltage Selector. Make sure that the label visible in the win-
dow indicates the correct voltage (i.e., the voltage of your mains).
To replace the fuses and/or change the selected voltage, see paragraph 12.1-Electrical.
4.3.3
Mains Cord
It is a standard cable, Cat. # E-WP008. Make sure your power outtake is provided with a reliable
ground connection, see also 4.4 & 4.5.
4.4
Connections
Connect the mains cord to a power outtake,
provided with a reliable earth connection
.
Beside the power module described in paragraph 4.3, the BP Recorder back panel encompasses
the following connectors, from left to right:-
Figure 10 “Back Panel”
SERIAL INTERFACE:
this delta 9-pin connector enables the operator to connect the in-
strument to the serial port RS232C of the PC, see paragraph 3-
DATA ACQUISITION.
PRESSURE SIGNAL
(ANALOGIC):
PULSE SIGNAL:
these connectors are useful to route both pressure and pulse
signals to CRT (oscilloscope), chart recorder or PC (via a data
acquisition system)