background image

 

UGO BASILE - I

TALY

 

Page 3 

Series 58000 Instruction Manual (Rev.1)

 

  

The piston motion reverses and the decompression phase begins. The systolic pressure 
is indicated, this time, by the return of the pulse trace.  

The pulse rate of the animal is automatically assessed in real time by a pulse rate 
counter which picks the signal from the pulse transducer.  

 

Figure 2 “Pneumatic Circuit” 

In both compression and decompression runs, even the 

diastolic, if selected, 

is re-

corded as the pressure at which the pulse shows an unmistakable transition from steady 
to decreasing amplitude.  

See Figure 3 “Sample of a Graph” and more details about graph and captions on para-
graph 7.3-Starting the Pressure Measurement. 

You will notice the pressure indications are sets of three digits, e.g., 070 – 130 – 180,  
marked on the lower edge of the recording chart in vertical pattern, for economy of 
space, as in the Chinese signs. 

Содержание 58500

Страница 1: ...UGO BASILE S R L sales ugobasile com service ugobasile com www ugobasile com instruction manual BP RECORDER Cat No 58500 58550 58600...

Страница 2: ...MUST BE FOLLOWED TO ENSURE SAFE OPERATION AND TO RETAIN THE INSTRUMENT IN SAFE CONDITIONS SERVICE AND ADJUSTMENTS SHOULD BE CARRIED OUT BY QUALIFIED PERSON NEL AUTHORIZED BY UGO BASILE ORGANIZATION AN...

Страница 3: ...al pressure of the artery systolic pressure on the paper grid on which the actual pressure of the system is digitally printed in 10 mm Hg steps At the end of the recording a second pressure measure me...

Страница 4: ...606 58000 343 1 1 1 3 mm PULSE PICK UP 6 mm PRESSURE CUFF WITH SET OF O RINGS SET OF 5 MEMBRANES MOUSE HOLDER 30 mm DIAM STANDARD SET FOR 58700 58000 518 58000 620 1 1 18 mm PULSE PICK UP 20 mm PRESSU...

Страница 5: ...CHECK 8 4 2 NOTES ON THE INSTRUCTION MANUAL 9 4 3 BEFORE APPLYING POWER 9 4 3 1 MAINS SWITCH 9 4 3 2 FUSE HOLDER VOLTAGE SELECTOR 10 4 3 3 MAINS CORD 10 4 4 CONNECTIONS 10 4 5 ADDITIONAL SAFETY CONSI...

Страница 6: ...9 3 2 ANIMAL GENDER 24 10 QUICK SETTING OF THE BP RECORDER 24 11 MEMORY MANAGEMENT 25 11 1 MEMORY RESET 26 11 1 1 MEMORY FULL MESSAGE 27 11 2 SETTING TIME DATE 27 11 3 DISPLAY CONTRAST 28 12 MAINTENA...

Страница 7: ...gure 2 Pneumatic Circuit 3 Figure 3 Sample of a Graph 4 Figure 4 Front Panel 4 Figure 5 Tail Cuff 6 Figure 6 Pulse Transducer 6 Figure 7 Rat Holder 7 Figure 8 Rat Scanner 8 Figure 9 Power Module 9 Fig...

Страница 8: ...analog output for direct connection to most digital recorders as PowerLab for data acquisition and management Its paramount features are i a set of pulse transducers of superior performance ii a relia...

Страница 9: ...ed via an appropriate piping 13 to an expansion vessel 7 a blow out valve 9 a pressure gauge 10 and the tail cuff 12 through the pneumatic connection 11 see Figure 2 Pneumatic Circuit The cuff pressur...

Страница 10: ...er Figure 2 Pneumatic Circuit In both compression and decompression runs even the diastolic if selected is re corded as the pressure at which the pulse shows an unmistakable transition from steady to...

Страница 11: ...o depressurize the system and to interrupt and abort the trial at any time whereupon the trial is brought to an end A mis take manoeuvre and the struggling of the animal are the most common reasons to...

Страница 12: ...ight and reassures the operator that the pressure generator is doing its job 2 3 Chart Recorder This high resolution 50mm real time thermal printer of GSI LUMONICS located on the left side of the uppe...

Страница 13: ...ng it or damaging the rubber membrane requires the cuff to be deflated To depressurize push the red key DECOMP this causes the solenoid valve 9 to open releasing the pneumatic system and the piston to...

Страница 14: ...ided with an ample U shaped tail slot e capability of dissipating the rat body heat by a suitable combination of ventilation slots selection of heat conductive materials etc Figure 7 Rat Holder See al...

Страница 15: ...communication between 58500 BP Recorder and an IBM or compatible PC is controlled by Ugo Basile Win DAS Software Cat 52010 supplied as optional See 13 5 Optional The 52010 is a Windows based Data Acq...

Страница 16: ...se the instrument Free of charge copies of the instruction manual are available upon request please con tact our service department see paragraph 12 5 Customer Support specifying the se ries number of...

Страница 17: ...Electrical 4 3 3 Mains Cord It is a standard cable Cat E WP008 Make sure your power outtake is provided with a reliable ground connection see also 4 4 4 5 4 4 Connections Connect the mains cord to a...

Страница 18: ...ECT INSTALLATION PROCEDURE INCORRECT OR IMPROPER USE OR IN ANY CASE NOT IN AC CORDANCE WITH THE PURPOSE FOR WHICH THE INSTRUMENT HAS BEEN DESIGNED AND THE WARNINGS BROUGHT TO YOUR ATTENTION IN THE INS...

Страница 19: ...g the BP Recorder has perfected the tech nique adding each his her personal touch to optimize technique for lab or application 5 2 Restraining the Rodent Basically rats and mice are placed into suitab...

Страница 20: ...succession You will notice a change only in the 3rd and 4th line which form a small type double string The top string shows in or der systolic diastolic pressure and heart rate in the com pression ph...

Страница 21: ...animal and its logic circuit thinks that something is wrong When you shake the transducer the dot enlarges to form a dash whose length is proportional to the amplitude of the signal To check the pneum...

Страница 22: ...t run you should select SPEED 1 of 10 mm Hg second or SPEED 2 of 20 mm Hg second The second speed refers to the chart motion in millimetres per second L low of 5 mm s SPEED 1 or 10 mm s SPEED 2 and H...

Страница 23: ...set the pressure value at which the reverse motion of the piston should start and confirm it by the F3 OK key If this function is not required depress F4 NUL 6 1 5 Set Reverse Time If you don t select...

Страница 24: ...ion RELIABLE SIGNAL Depress the blue key START REVERSE to begin the meas uring sequence see paragraph 7 3 Starting the Pressure Measurement A bar proportional to the amplitude of the signal appears ab...

Страница 25: ...ion is automatically adjusted to the selected speed SPEED 1 or SPEED 2 You will see from the display that the top line indicates the cuff increasing pressure up to dated at one second inter vals In th...

Страница 26: ...pulse transducer In the top line we read SYST UP and the pressure value 141 mm Hg in this example at which the pulse amplitudes is re duced to nil The vertical line represents the division between th...

Страница 27: ...orded see the example taken from a real experiment where we read from the top of the data set date time systolic pressure compression run diastolic pressure compression run systolic pressure decompres...

Страница 28: ...ion in the piping cylinder and expansion vessel plus the mercury oscilla tions in case a mercury gauge is used As soon F2 STP is depressed this display will appear In case of discrepancy between displ...

Страница 29: ...nter the PATH menu and press the F4 ESC key selecting no option 9 2 Instrument Address and Baud Rate When the SER or ALL functions are activated and the 585600 is correctly connected to the PC the mes...

Страница 30: ...decrease of one step By keeping them pressed longer the selection can be made at higher speed The selected features will be kept in the instrument s memory even when it is turned off To erase one or m...

Страница 31: ...HE BP RECORDER This Instruction Manual explains extensively how to take advantage of all the features of fered by this microprocessor controlled unit The following table summarizes features however we...

Страница 32: ...lay shows all incoming data in real time 11 MEMORY MANAGEMENT Each trial no matter whether printed or sent to the PC is saved in the memory which can store about 300 strings in cluding the result of t...

Страница 33: ...9 25 13 S163D070p133 gMa03 2003 09 05 09 27 58 s169d077p133 gMa04 2003 09 05 09 32 18 S161D070p133 gMa04 2003 09 05 09 36 12 Capital S and D mean systolic and diastolic in decom pression phase Small s...

Страница 34: ...previously explained see paragraph 11 MEMORY MANAGE MENT 3 you may also ignore the indication of MEMORY FULL and go on with the experi ment by depressing the F4 ESC key The data acquired onwards will...

Страница 35: ...ssed until you reach the desired display contrast The readability of the liquid crystal display depends on the angle of view select the ideal contrast according to the height of the table the operator...

Страница 36: ...ed Check the voltage flag visible in the window before applying electrical power 12 2 Replacing a Leaky Pressure Cuff Membrane Take out the leaky membrane Fit the fresh membrane into the cuff tube and...

Страница 37: ...hould be stored away from daylight which in the long run degrades their con sistency Calibrate instrument before using see paragraph 8 CALIBRATION 12 5 Customer Support For any further information you...

Страница 38: ...mm pulse pick up 6 mm pressure cuff and mouse holder 58700 BP RECORDER FOR DOG as above but complete with accessories for Dog 18 mm pulse pick up and 20 mm pressure cuff All models are complete with f...

Страница 39: ...mm Cuff 58000 709 Set of 5 Membranes for 9 mm Cuff 58000 713 Set of 5 Membranes for 13 mm Cuff 58000 720 Set of 5 Membranes for 20 mm Cuff 58000 724 Set of 5 Membranes for 24 mm Cuff 58000 806 Set of...

Страница 40: ...x Operating Temperature 0 to 40 C Sound Level 70 dB Weight net Kg 10 6 Shipping Weight Kg 15 0 approx Dimensions 35 x 35 x 16 5 h cm Packing dimensions 67 x 42 x 53 cm 14 BIBLIOGRAPHY M Gerold H Tschi...

Страница 41: ......

Страница 42: ...1EL01 sh 5 of 5 Electronic List 58000 001EC01 Board Component Layout 58000 001 ES02 Wiring Diagram 58000 001 EL02 Electronic List 58000 001 EC02 Board Component Layout 58000 001 EW01 Cables Connectors...

Страница 43: ......

Страница 44: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Страница 45: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Страница 46: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Страница 47: ...actured in compliance with the following European Union Directives and relevant harmonized standards 2006 42 CE on machinery 2014 35 UE relating to electrical equipment designed for use within certain...

Отзывы: