Page 25
37360 Instruction Manual (Rev. 0)
Calibration is recommended, if the Radiometer 37300 is available, see paragraph 9-
CALIBRATION.
10.5
Customer Support
For any further information you may desire concerning the use and/or maintenance of the
Tail Flick Unit and accessories, please do not hesitate to contact our
service department
(or our local distributor) either directly of via our support page
http://www.ugobasile.com/support.html
:
UGO BASILE s.r.l.
Via G. Di Vittorio 2
21036 GEMONIO – Varese, ITALY
Phone : +39 0332 744574
[email protected]
[email protected]
[email protected]
Before sending any instrument to our factory for repair
, please contact our logistics
department to obtain a return authorization number (RMA) and shipping/packing instruc-
tions.
We may not be held responsible for damages during transport due to poor packing;
whenever possible, please use the original packing.
11 INSTRUMENT
SPECIFICATIONS
Operation
Power Requirement
Universal mains 85-264 VAC, 50-60Hz
Commands
via
soft-buttons
Read-out
multifunction graphic display
Print-out Via
optional
thermal MiniPrinter or panel-mount printer
Starting
via pedal switch. Additional key on the Controller
I.R. Intensity
Adjustable in the interval 01-99 (in one digit steps)
Latency Time
displayed on the graphic display, in 0.1s steps
Connection to PC
via USB
Data Portability to PC
Via USB pen-drive provided
Calibration
Via appropriate I.R. Radiometer, see paragraph 9.1
Operating Temperature
10° to 40° C