37360 Instruction Manual (Rev. 0)
This command can be useful, for instance, before starting a new experimental session, if
the operators prefers to cancel the data saved from previous experiments.
The Memory menu automatically appears also when the memory is
fully loaded
, which
means it has reached its maximum capability of about 300 measures.
7.1.1
“Memory Full” Message
If the “
MEMORY FULL
” message appears during acquisition, the operator has three op-
tions:
if not interested in the data
saved in the memory, the re-
searcher may simply depress the
F4 (ESC)
key, followed by the
F3
(RES)
and confirm the reset by
answering
F3 (YES)
to the ques-
tion “
ARE YOU SURE
?”, see para-
graph 7.1.
When
the data
previously saved in the memory are
to not to be cancelled
,
the researcher may print and/or send them to the memory key or PC as previ-
ously explained (see paragraph 7-MEMORY MANAGEMENT).
There is also the possibility to
ignore the indication of “
MEMORY FULL
”
and
go on with the experiment. The data acquired onwards will not be saved.
Consider that, in this case, the memory remains fully loaded and the reminds it
by showing the message “
MEMORY FULL
” at each measure.
7.2
Setting Time & Date
To set time and date, from the Main menu depress the
F2 (FNC)
key, followed again by
the
F2 (CLK)
key.
Use the
F3
horizontal arrow key to highlight
the parameter to be modified.
Now use the
F1 (UP )
or
F2 (DWN)
keys to
set the correct value; a momentarily action
on the keys will increase or decrease the
numerical value of one step: depress the
keys longer for faster selection.
Depress the
F2
(
ENT
) key to confirm the selection.