SPECIFICATIONS
Parameter
Range
Resolution
Accuracy
Temperature Measurement
Flue Temperature
Inlet Temperature (Internal Sensor)
Inlet Temperature (External Sensor)
32˚ to 1112˚F (0˚ to 600˚)
32˚ to 122˚F (0 to 50°C)
32˚ to 1112˚F (0˚ to 600˚)
0.1˚F (0.1˚C)
0.1˚F (0.1˚C)
0.2˚ (0.1˚C)
±(0.3% rdg +3.6˚F (2˚C))
±2˚F (1˚C)
1˚F (0.5˚C)
Flue Gas Measurement
Carbon Monoxide
Carbon Dioxide
0 - 2000ppm
0 - 20%
1ppm
0.1%
±3ppm or ±5% of Reading
(whichever is greater)
±0.3% Volume
Calculations
Oxygen
CO/CO
2
Ratio
Efficiency (Nett or Gross)
Efficiency High (C)
Excess Air
0 - 21%
0 - 0.9999
0 - 99.9%
0 - 119.9%
0 - 119.9%
0.1%
0.0001
0.1%
0.1%
0.1%
±0.3% Volume
±5% of Reading
±1% of Reading
±1% of Reading
±0.2% of Reading
Pre-programmed Fuels
Pellets, Light Oil, LPG, Butane, Propane, Natural Gas, Bio Oil, Heavy Oil
Battery Life
>8 hours (continuous with pump on)
Certification
The C161 is TUV-tested and certified to EN 50379, Parts 1-3 in accordance to 1st German Feder-
al Emission Control Ordinance (Bim5chV)
Operating Conditions
Temperatures
Humidity
32˚ to 113˚F (0˚ to 45˚C)
15 to 90% RH, (non-condensing)
Power Supply
Rechargeable batteries, USB Charging
Physical Characteristics
Weight
Dimensions
1.5 lb. (635g)
H: 8.54”x W: 4.18”x D: 1.86” (H: 217 mm x W:106 mm x D: 47mm)
The C161 is in conformity with the relevant Union harmonization legislation listed below:
Directive
Title
201430EU
Electromagnetic Compatibility (EMC)
201165EU
Restriction of the use of certain hazardous substances in electrical and electronic equipment (RoHS)
The following harmonized standards and technical specifications have been applied:
CERTIFICATION
The UEi C161 is TUV-tested and certified to EN 50379, Parts 1, 2 & 3 in accordance to 1st German Federal Emission Control Ordinance
(BlmSchV). Manufactured to meet AHRI 1260 standard.
EMC
EN507270:2015
Safety
EN61010-1:2010
RoHS
IEC62321-2:2013, IEC62321-1:2013; IEC62321-3-1;2013, IEC63321-5:2013, IEC623321-4:2013, IEC62321-7-2:2017, IEC62321-7-1:2015,
IEC62321-6:2015
UK CA