1. Error messages
Error
Possible cause
Remedy
Display Err - 300 Excessive cuff pressure.
The cuff was inflated to
its maximum due to the
movement of the arm
or body.
The air hose may be
kinked or the air plug
blocked.
- Repeat the measurement
- Do not move your arm
- Do not speak
Inspect the hose line and air
plug for blockages
Display Err - 1
Measurement error;
measurement could not
be carried out due to
movement or speaking.
- Repeat the measurement
- Do not move your arm
- Do not speak
The acquired values lie
outside the measuring
range of the device (see
„Technical data“ on
page 60).
Repeat the measurement
Display Err - 2
Inflation error; pressure
built up too quickly or
too slowly.
Check that the air connector
is fitted properly, repeat
measurement.
Fault has been iden-
tified. The cuff is not
correctly attached
- Repeat the measurement
- Do not move your arm
- Do not speak
Display Err - 3
Deflation rate too high
or too low.
Call the Service hotline if
this occurs repeatedly.
Display 0
0
Start/Stop button was
accidentally pressed
while the batteries were
being replaced or while
the mains adapter was
being plugged in.
Switch unit off and then on
again using the Start/Stop
button.
Arrhythmia, movement
disorders, trembling,
wobbling (objects), deep
breathing etc.
Repeat measurement after
resting for 3-5 minutes.
Listen out for the device‘s
pulse beep signals. Possible
influences on the measure-
ment results (depending on
the severity of the arrhyth-
mia) must be discussed with
your doctor!
Error
Possible cause
Remedy
The motion control
detected movement du-
ring the measurement.
- Repeat the measurement
- Do not move your arm
- Do not speak
Measured
values/pulse
values are
extremely high
or low.
Too little rest before ta-
king the measurement.
Movement or talking
during the measure-
ment.
Repeat the measurement
after a break of 3-5 minutes.
Pay attention to the usage
information on page 45.
Incorrect measurement
posture
Repeat the measurement
and maintain the correct
posture. (Page 49)
Measurements
taken in direct
succession
produce diffe-
rent measured
values.
Blood pressure is not a
fixed value. It can go up
or down by more than
20 mmHg within a few
minutes.
In order to obtain compara-
ble results, always carry out
the measurements under
the same conditions.
Measurement
was interrupted,
the cuff deflated
to 0 and the
measurement
restarted.
Device detects a fault or
that the diastolic value
is too low.
Device shows no fault.
Repeat measurement
Movement during mea-
surement
If necessary, interrupt mea-
surement and repeat after 5
minutes‘ rest.
No display
after device is
switched on
Batteries inserted
incorrectly
Check position of batteries.
Batteries flat
Replace batteries.
Battery contacts dirty.
Clean battery contacts with
a dry cloth.
Dashes in
display after
pressing Memo-
ry button
No measured values
have been stored.
Repeat measurement
Measurement is
interrupted
Batteries flat
Replace batteries.
The time is
visible in the
display, but the
measurement is
interrupted and
the symbol
appears.
Batteries flat
Replace batteries.