English
English
EN-54
EN-55
Operating the device
Operating the device
C
C
A more frequent appearance of the symbol may indicate cardiac
arrhythmia and must be discussed with your doctor! Measure-
ment results accompanied by irregular pulses should therefore
be viewed as critical and repeated under more favourable con-
ditions.
The symbol is stored with the respective measurement.
The device stores the measured values in two memories (one per
user). Each memory can store up to 120 readings and the aver-
age value (A). The average value is calculated from the last three
measurements taken.
The measured results are automatically saved in the memory. If
more than 120 measured values are saved, the oldest measured
value (no. 120) is deleted so that the latest value (no. 1) can be
recorded.
To retrieve data, press the Memory button of the user you wish to
retrieve the data for while the device is switched off. The average
value is displayed with „A“ (Figure 1).
Set the user whose data you wish to retrieve.
The symbol is stored with the respective measurement.
Data retrieval
Movements during the measurement may re-
sult in incorrect measurement values. The mo-
tion control registers stronger movements and
these are indicated by the symbol in the dis-
play. Other disruptive factors such as speaking,
coughing or cuff noises can cause the symbol to
be displayed. If this symbol appears, please repeat the measure-
ment and avoid the disruptive factors.
15. Using the memory
14. Motion control
Figure 2
Figure 3
Figure 4
Press the Memory button again to display the
most recently measured values (Figure 2). If the
date/time function is activated, the memory
number (Figure 2), the date (Figure 3) and the
time (Figure 4) are displayed alternately.
Figure 1
Press the Memory button repeatedly to recall further measured values.
You can switch between the two memories by pressing Memory
button 1 or 2.
The stored data remain visible in the display for around 30 sec-
onds. The device then switches itself off.
To delete individual measured values, recall the
measured value you wish to delete.
Press the Memory button again and hold down
until the display firstly flashes and is then re-
placed with dashes.
Deleting data
To delete the entire memory, retrieve the re-
spective average value (A), press the Memory
button again and hold down until the average
value firstly flashes and is then replaced with
dashes.