
2
IM_ONE-01
www.ueonline.com
MOUNTING
Tools Required
1-1/16" wrench for sensor hex fitting
Screwdriver for mounting bolts
2 mounting bolts (1/4" Max.)
BEFORE INSTALLING, CHECK THE SENSOR MODEL SELECTED FOR COMPATIBILITY TO THE PROCESS MEDIA IN CONTACT WITH THE SENSOR AND WETTED
PARTS.
FOR PRESSURE AND LOCAL TEMPERATURE MODELS ALWAYS HOLD A WRENCH ON THE SENSOR HEX WHEN MOUNTING UNIT. DO NOT TIGHTEN BY TURNING
ENCLOSURE, THIS WILL DAMAGE THE CONNECTION BETWEEN THE SENSOR AND HOUSING.
INSTALL UNITS WHERE SHOCK, VIBRATION AND TEMPERATURE FLUCTUATIONS ARE MINIMAL. MOUNT UNIT TO PREVENT MOISTURE FROM ENTERING
ENCLOSURE. USE PROPERLY RATED SEALING FITTINGS FOR ELECTRICAL WIRE ENTRY. DO NOT MOUNT UNIT IN AMBIENT TEMPERATURES EXCEEDING
PUBLISHED LIMITS. THIS IS ESPECIALLY CRITICAL FOR LOCAL MOUNT TEMPERATURE UNITS.
IN HIGH SHOCK AND VIBRATION APPLICATIONS, SECURE THE ENCLOSURE AS DETAILED ON THE NEXT PAGE. DO NOT MOUNT VIA THE PROCESS
CONNECTION ONLY.
FOR DIFFERENTIAL PRESSURE MODELS (ESPECIALLY LOW RANGE UNITS), CARE SHOULD BE TAKEN TO MOUNT THE SENSOR LEVEL TO MINIMIZE ANY
PRESSURE READING OFFSETS. THE OFFSET COMMAND MAY BE USED TO ZERO THE DISPLAY, SEE PAGE 8 FOR ADDITIONAL INFORMATION.
Electronics Housing
Mount the unit using the two (2) 1/4" clearance holes in the mounting ears. Plumb sensor to the process port. See page 14 for dimensions.
The One Series product may be mounted in any position except with the electrical conduit connection facing up. Ensure the process connection is sealed to the
process port to prevent leakage. Care should be taken to minimize effects of shock and vibration.
PROCESS CONNECTIONS AND SENSOR INSTALLATION
Pressure and Differential Pressure Models
To pipe mount: Thread the pressure connection onto the pressure port, making sure that the mating threads are clean and free of debris. Use a wrench on the
hex pressure connection to tighten. Test for leaks. On Differential Pressure models, the Low (L) side pressure must NOT exceed the high (H) side pressure.
WARNING: NEVER INSERT ANY OBJECT INTO THE PRESSURE SENSOR OPENING. DAMAGE TO THE SENSOR WILL RESULT, AFFECTING ACCURACY.
Local and Remote Temperature Models
For Remote Sensing (models C, H & R): Route the extension wire to avoid contact with live components. Avoid kinks, or excessive flexing.
For Surface Sensing (models C, H & R): Secure the sensor housing to the pipe or vessel using an adhesive or strapping method suitable for the application.
For Immersion Sensing (models C, H, R & L): Use of a thermowell is highly recommended to aid in maintenance, testing and preservation of the system integrity.
Insert the sensor housing (0.25" diameter) into the well ensuring that the housing bottoms out and will be completely immersed in the media (2.5" min.) Secure
the sensor using an appropriately sized compression fitting (PF73 or equivalent) or union connector (SA6213-348 or equivalent.)
For Local Ambient Sensing (model L): Mount using the mounting ears on the electronics housing. Mount the product to ensure that the sensor housing will not
be damaged and where the measured temperature is representative of the surrounding environment.
For best temperature measurements, the sensor housing must be in full contact with the surface or media being measured. Heat transfer compound may be
used to aid in fully transferring the media temperature to the sensor housing. Locate where the temperature is most representative of the system. Minimum
insertion depth is 2-1/2".
CLEANING THE FACEPLATE
One Series models with ATEX approvals have a urethane based conductive coating applied to the faceplate perimeter. The cleaning
of this surface should be performed with a damp cloth only. Should the surface exhibit signs of flaking or wear, the product should
be removed from service in hazardous locations where ATEX approvals are required.
Содержание 2W2D
Страница 13: ...IM_ONE 01 13 www ueonline com PROCESS DISPLAY ...