kANGAROO
™ 924
8
Enteral Feeding Pump
• A mild detergent should be used for general cleaning. If necessary, the pump may be cleaned with a 10:1 water and
hypochlorite mixture, however, repeated cleaning with this solution can damage the plastic pump case. Isopropyl alcohol
applied with a damp (not wet) cotton swab may be used for cleaning difficult-to-reach areas, such as the drop sensor and
inside the A.C. receptacle, however, it should be used sparingly because repeat cleaning may damage the case.
• DO NOT USE strong cleaners such as Spray-Nine™*, Phisohex™*, Hibiclens™*, or Vesta-Syde™* because damage to the pump
case housing can result.
Directions For Cleaning Pump Housing
• Refer to General Cleaning Directions before starting.
• Clean outside surface with a damp (not wet) cloth or sponge and keep pump in upright position as much as possible.
• Avoid excess moisture near handle hinges.
• Avoid excess moisture near pole clamp area.
Directions For Cleaning Pump Power Cord
• Refer to General Cleaning Directions before starting.
• Unless soiling is observed, the power cord should not be cleaned.
• If cleaning of the power cord is necessary, wipe the exterior surfaces of the wall plug of the power cord with a cloth dampened
with isopropyl alcohol.
CAUTION Avoid exposing power cord to excess moisture, as this can lead to an electrical shock or fire hazard.
Directions For Cleaning Drop Detector
• Refer to General Cleaning Directions before starting.
• Clean the drop sensors in the upper drip chamber guide with a cotton swab dampened with isopropyl alcohol.
• Be sure sensor areas are clear.
Directions For Cleaning OPTIX Lock
• Refer to General Cleaning Directions before starting.
• Clean lens in the OPTIX lock area with a cotton swab dampened with isopropyl alcohol.
• Be sure sensor areas of the lens are clear.
• Avoid using abrasive cleaners to prevent scratching/damaging the lens.
Directions For Cleaning Rotor Assembly
• Refer to General Cleaning Directions before starting.
• Loosen rotor set screw with a 5/64" (2MM) allen wrench and gently pull rotor forward off shaft. After removing rotor, avoid
getting any moisture in the rotor shaft opening.
• Clean rollers thoroughly with warm soapy water, or isopropyl alcohol if necessary.
• Be sure all parts of rotor are completely dry before putting it back onto shaft.
• To replace rotor, align set screw on hub of rotor with the flattened section of the output shaft.
• Push rotor into place and tighten set screw (Do not over tighten).
Status: Review Routing On: CR10000186 Doc#: AW10001678
11/10/2006 Artwork#s: WN 182493-01 Job#: 14761