12. Grasp silicone tubing and gently stretch around rollers (See Figure 5A). Insert OPTIX connector into OPTIX lock (B). Thread
tubing over tubing guide (C). Secure latch over tubing (D) (See Figure 5B). Do not over-stretch silicone tube.
VERIFY THAT THE PUMP SET IS PROPERLY LOADED IN THE PUMP BEFORE PROCEEDING. AN IMPROPERLY LOADED PUMP SET
COULD RESULT IN UNCONTROLLED FLOW AND POSSIBLE PATIENT INJURY.
13. Set Volume-To-Be-Delivered, if desired. Press VTBD, then rotate adjustment knob or press or to set VTBD amount.
14. Open control clamp.
15. Press START/HOLD.
16. Proceed with feeding.
17. Dispose of pump sets (Not to exceed 24 hours of use).
Used pump sets shall be handled as follows:
To prevent employee exposure risk, the materials must be handled in accordance with bloodborne pathogen standards before being
placed in an approved disposal container.
Disposal must be performed in accordance with current medical practices or local regulations in regards to disposal of infectious,
biological or medical waste.
Example of proper placement of Pump Set
in KANGAROO 924 Feeding Pump.
Note: When in use with an IV pole, assure feeding bag
is set so that the top of the fluid column is 28 inches
above the top of the pump.
Figure 6