kANGAROO
™ 924
4
Enteral Feeding Pump
Battery Operation
Unplugging the pump from the AC line outlet will automatically put the pump on battery power. A new battery, when fully charged,
will operate the instrument for more than 16 hours at 125 milliliters per hour. Approximately 15 minutes prior to battery discharge, a
low battery alarm will occur, (see alarm section). When complete discharge occurs, the pump will automatically turn itself off.
NOTE: AS THE BATTERIES GET OLDER, THE TIME FROM LOW BATTERY ALARM TO COMPLETE DISCHARGE MAY BE LESS THAN 15 MINUTES.
To recharge battery, insert power cord into AC wall outlet. Battery will automatically begin charging. The recharge time for a fully
discharged battery is approximately 12 hours.
The battery will charge whenever the pump is plugged into a wall outlet, even when the OFF button has been pressed. To be certain that
power to all circuits has been disconnected, the power cord must be unplugged from the wall outlet.
Battery
Replacement battery packs are available from Tyco Healthcare Group LP (See Section VI - Factory Service)
WARNING Replace the battery pack only with the same type and rating of battery.
Beginning Administration
Below are general instructions for inserting all Tyco Healthcare Group LP Pump Sets into the KANGAROO 924 Enteral Feeding Pump.
1. Press
ON
.
2. Wait while pump performs system check. (Displays “
“ for a moment.)
3. Check
VOL CLR
light or press
VOL
to insure that volume delivered is clear, if desired.
4. Rate of delivery may be changed when the delivery rate is shown on the display by using the adjustment knob or by
pressing or button. Holding button in will cause numbers to continue to increase or decrease.
A
Figure 4A
Figure 4B
5. Slide drip chamber into upper drip chamber guide (See Figure 4A).
6. Seat and secure chamber in lower drip chamber guide (A).
7. Remove distal connector protective cover from the feeding set..
8. Follow 924 feeding set instructions to prepare for feeding.
9. Slowly open control clamp and fill entire line with fluid, leaving a few drops of fluid in the drip chamber.
NOTE: AVOID FILLING THE DRIP CHAMBER MORE THAN HALF FULL. Close control clamp.
10. Attach distal connector to feeding tube.
11. Inspect lens in OPTIX lock area (B). Clean any contaminants from lens before proceeding to the next step (Figure 5B).