kANGAROO
™ 924
7
Enteral Feeding Pump
VTBD Delivered (Vtbd dEL)
Preset VTBD has been delivered to patient.
When preset VTBD has been delivered, perform one of the following:
– Repeat the VTBD amount by pressing START/HOLD after clearing the alarm.
– Cancel the VTBD feature by pressing and holding VTBD for 5 seconds.
– To feed additional volume, increase VTBD amount by first pressing VTBD. Then rotate knob or press until
new VTBD is displayed.
– Turn pump off.
NOTE: FOR ALL ALARM CONDITIONS, PRESSING THE START/HOLD BUTTON WILL SILENCE THE AUDIBLE ALARM. THE
VISIBLE DISPLAY WILL CONTINUE TO FLASH UNTIL THE START/HOLD BUTTON IS PRESSED AGAIN.
The alarm volume is set using the front panel in one of two ways:
1. Using the “Biotech Mode” as described. (See page 10)
2. While the alarm is sounding, press to increase the volume, press
to decrease the volume, or rotate knob.
The volume is changed in steps from 1 to 10, with 1 the quietest and 10 the loudest.
WARNING If alarm does not sound during system check, or when any alarm condition exists, see “No Audible
Alarm” symptom in Troubleshooting guide.
Section IV Cleaning
Cleaning is to be performed as required. The user will define the cleaning intervals based on their knowledge of the environment in
which the pump is used. Cleaning may become necessary between intervals due to unforeseen soiling. Only personnel trained in the
cleaning of medical devices shall perform cleanings.
Important User Information
CAUTION DO NOT IMMERSE PUMP OR POWER CORD IN WATER OR OTHER CLEANING SOLUTION; CLEAN USING A DAMP
(NOT WET) CLOTH OR SPONGE. FAILURE TO FOLLOW THE CLEANING PROCEDURES DESCRIBED HEREIN COULD
RESULT IN HAZARDS TO USERS. As with any A.C. powered electrical device, care must be taken to prevent
liquid from entering the pump or the power cord receptacle to avoid electrical shock hazard, fire
hazard, or damage to electrical components.
Should any of the following events occur, DO NOT USE the pump until it has been properly cleaned and serviced by personnel
trained in servicing KANGAROO pumps:
• wetting of the pump’s power cord or leakage into the pump interior during cleaning.
• spillage of large amounts of formula onto the pump exterior or any spillage on the power cord.
General Cleaning Directions
Cleaning of KANGAROO 924 pumps must be performed as follows:
CAUTION Disconnect pump from A.C. power source before cleaning. Upon completion of cleaning, do not connect
to A.C. power source until pump and power cord are thoroughly dry.