kANGAROO
™ 924
9
Enteral Feeding Pump
Preventive Maintenance
This pump must be periodically serviced to assure proper functioning and safety. The recommended service interval is six (6) months, but
in no case less than once each twelve (12) months per JCAHO. Servicing must be performed by personnel trained in servicing KANGAROO
pumps and may be done at the user’s biomedical engineering department, an outside service, or by Tyco Healthcare Group LP Factory
Service. To arrange for Tyco Healthcare Group LP Factory Service, call 1-800-448-0190 in the US or contact your local customer service
center.
The KANGAROO 924 pump contains no user serviceable parts. User maintenance consists of cleaning the exterior of the pump as stated in
the cleaning section of this manual. All other maintenance is to be performed by the user’s appropriately qualified technical personnel.
Section V Troubleshooting Guide
Listed below are some of the probable causes of alarm conditions and their corrections.
Flow Error Alarm
(FLo Err)
Feeding container is empty.
– Refill feeding container. If bag has been in use for 24 hours, replace
Feeding tube or feeding set tubing is occluded.
– Locate point of occlusion and correct. Note: occlusion may occur upstream or downstream from pump.
• Feeding Container
• Feeding Pump Set (control clamp closed)
• Feeding Tube - patency of feeding tube should be checked.
Drip chamber is improperly placed in pump.
– Check to make sure drip chamber is properly “locked” in position in drip chamber guide.
Drip chamber walls are coated with feeding formula.
– Check to make sure formula is not preventing detectors from proper operation. If formula cannot be removed from inside
walls of chamber via chamber manipulation, replace set.
Sensors in upper drip chamber guide are blocked.
– Check to make sure detectors are free of dried formula. Remove any deposits by using a cotton swab dampened
with warm soapy water.
Low Battery Alarm
(Lo bAT)
Battery has been run down below point of maintaining accuracy of pump.
– Plug pump power cord into wall outlet.
Hold Error Alarm
(HLd Err)
Pump has been left in the hold mode for over 5 minutes.
– Press START/HOLD to silence alarm, then press START/HOLD again to clear the visual alarm message.
Press START/HOLD again to restart.