7.4. Other Problems and Solutions
1. The display of the flowmeter measures normal "R", and the received signal
strength and signal goodness are good. The fluid in the pipeline under test has
been flowing, but the flow rate of the flowmeter has been shown to be 0.0000 at
this time, what is the reason?
A: It is possible that the user used the "static zero" operation in the presence of
fluid flow.The solution is to go to window M43 to "clear static zero zero" and select
"YES".
2. What is the reason for the flow data displayed by the flowmeter to be smaller or
larger than that of the pipeline?
A :(1) Incorrect Settings were made in the M44 window;The solution is to enter the
M44 window and enter the value "0".
(2) The sensor is not installed correctly.
(3) There is a zero point in the flow meter. Under the condition of confirming that
the fluid in the pipeline is completely stationary, it enters into the M42 window for
"static zero setting" operation.