Chapter 4 Flow measurement
- 43 -
the maximum signal strength, and also check that the ratio of measured and
theoretical transmission time is between 97% and 103%.
4.4.2. Transfer time ratio
The M91 window is used to display the transmission time ratio, which is the
percentage value of the actual measured time and the transmission time calculated
according to the parameters set by the flow meter.If this value exceeds 97%~103%,
it indicates that either the parameter setting is wrong or the installation distance is
wrong,please check separately.
4.5. Measurement data view
When the signal strength and signal quality, transfer time than meet the
requirement of measurement, the measurement data obtained is accurate, can see
the measurement in 00 ~ 09 menu data, including the instantaneous flow rate and
the instantaneous velocity, the cumulative amount of positive and negative,
cumulative amount, date, time, heat flow, total quantity of heat, temperature,
current value and the corresponding value of the analog input, net cumulative flow
today, and so on.
4.6. Measurement data processing
After the ultrasonic flowmeter (energy meter) is measured normally, the measured
data obtained can be printed in real time or timed, and can also be stored in the
built-in memory.The data in the memory can be processed by purchasing flow data
statistics and analysis software.For data storage methods and usage methods of
traffic data statistical analysis software, please refer to Chapter 5 Data Processing.
Содержание MC-SUP2000H
Страница 1: ...Medidor de flujo ultrasónico en líquidos portátil Meacon MC SUP2000H ...
Страница 5: ......
Страница 11: ...Chapter 1 Introduction 6 1 6 Parts Identification 1 6 1 Converter Front view Side view ...
Страница 16: ...Chapter 1 Introduction 11 View and size Small size HS Medium size HM ...
Страница 17: ...Chapter 1 Introduction 12 High temp Small size HS HT ...
Страница 18: ...Chapter 1 Introduction 13 High temp Medium size HM HT Extension bracket ...
Страница 69: ......