Turtle Tough TT-TOUGH-DO-M4M Скачать руководство пользователя страница 9

TURTLE

TOUGH

®

 

9

Calibration

Calibration Verification Check

Every instrument leaves our manufacturing facility in a calibrated condition after going through a test programme 
of several hours. The user should nevertheless re-check the calibration since a change may have occurred due to 
long transport or storage periods.

For a fast routine check it is sufficient to have the sensor in atmospheric air. The air calibration table on the front 
panel of your instrument shows the appropriate oxygen calibration value per degree Celsius.

For example:

 20 °C equals 9.7 mg/l O₂ or 21 °C equals 9.5 mg/l.

• 

If the liquid-crystal display of the instrument indicates, for example a temperature of 20.5°C, the changeover 
to oxygen should produce the reading 9.6 mg/l O₂. This is the arithmetic mean between the values 20 °C 
(9.7 mg/l O₂) and 21 °C (9.5 mg/l O₂).

• 

If the value displayed does not correspond to the value on the table or to the calculated value, it will be 
necessary to recalibrate using the calibration potentiometer by means of the screwdriver included.

• 

If the setting is correct, the device is ready for carrying out measurement.

Calibration in Atmospheric Air

The 1-point calibration in atmospheric ambient air is the simplest type of calibration. The front panel of the 

 

instrument incorporates the calibration table for air calibration.

1.  Put the sensor in a shaded place (avoid direct sunlight). The tip of the sensor and the membrane have to be 

dry - remove any humidity by dabbing the tip with tissue paper. Cover the sensor i.e. with newspaper to 
avoid any wind on the tip.

2.  Wait for at least 15-20 minutes to be sure that the sensor has reached ambient temperature.
3.  After determining the air temperature please read the associated calibration value on the air calibration table. 

Switch over to O₂ and calibrate using the calibration potentiometer by means of the screwdriver included.

Note:

 For high accuracy calibration of your instrument, barometric pressure should be taken into consideration.

Calibration to Barometric Pressure

Different air pressures result in different calibration values, so for high accuracy readings a calibration of your 
instrument taking barometric pressure into consideration must be performed. If you know the pressure for your 
location you can correct the calibration value according to the following formula:

 

Corrected calibration value [ppm] = Pressure" [mbar ] x Table calibration value [ppm] / 1013 [mbar]

This is clarified further by the following example:

You have measured a temperature of 20 °C. The air calibration table states the relevant calibration value of 9.7 ppm.
If the environment has a different pressure than 1013 mbar, the calibration value has to be corrected. For example 
if we are at 1500m above sea level then the pressure per the pressure table (shown above) equates to 838 mbar. 
Then the corrected calibration value would be as follows:

838 x 9.7 /1013 = 8.0

After determining the correct calibration value, set the calibration potentiometer accordingly by means of the 
screwdriver included.

Содержание TT-TOUGH-DO-M4M

Страница 1: ...TURTLE TOUGH 1 TURTLE TOUGH turtletoughsensors com MANUAL M4M TOUGH DISSOLVED OXYGEN SYSTEM Dissolved Oxygen...

Страница 2: ...emicals 5 Sensors 5 Introduction 6 Specifications 7 Unpacking 7 Operation Temperature Measurement 8 Temperature Stabilisation Time 8 Dissolved Oxygen Measurement 8 Calibration Calibration Verification...

Страница 3: ...eading Safety Precautions the special safety precautions outlined in other sections must also be observed Warnings used in this manual This manual contains vital information relating to the safety of...

Страница 4: ...wired electrical outlet It is recommended that the outlet is protected by an Residual Current Detector or is in any event in compliance with all local electrical regulations Do not bury electrical su...

Страница 5: ...them CAUTION When handling this equipment the accident prevention regulations applicable on site must be observed and the specified personal protective equipment worn PPE examples of protective cloth...

Страница 6: ...d for zero solutions The sensor has an extremely fast response which is critical for applications that demand accuracy or measure at high ppm This is especially true for portable units which are often...

Страница 7: ...atmosphere Housing IP 65 PVC housing Dimensions 200 x 270 x 250 mm H x W x D approx Weight 2 3 kg approx Field case 385 x 355 x 180 mm H x W x D approx Weight 1 5 kg approx Unpacking Please have a cop...

Страница 8: ...antly kept in the measuring liquid it will not be necessary to observe a temperature stabilisation time The best procedure is to wait until the temperature value on the liquid crystal display remains...

Страница 9: ...nd the membrane have to be dry remove any humidity by dabbing the tip with tissue paper Cover the sensor i e with newspaper to avoid any wind on the tip 2 Wait for at least 15 20 minutes to be sure th...

Страница 10: ...material It is held by means of two red collars 4 and one O Ring 5 During measurements the membrane must not come into contact with sharp objects such as stones or sharp grained sand which might dama...

Страница 11: ...verified by holding the sensor horizontally You should now be able to see an air bubble no larger than 4 mm in the centre of the pressure compensation membrane Should refilling of the electrolyte be r...

Страница 12: ...low since the white wick like insert would otherwise be damaged Doing this voids the warranty 6 Wipe the cathode clean with tissue paper but do not polish it because the scratched surface helps to for...

Страница 13: ...ane until electrolyte appears in all the electrolyte channels Note Never use a needle or other sharp objects to create flow since the white wick like insert would otherwise be damaged Doing this voids...

Страница 14: ...pressure ring nut and screw it on carefully Using the Membrane Positioner Use the Membrane Positioner TT TOUGH DO MP to easily replace the cathode membrane without any wrinkles Follow these steps for...

Страница 15: ...Kit for sensor tip with 10 membranes 3 O Rings and 1 bottle electrolyte solution TT TOUGH DO MEM Refill Kit without electrolyte TT TOUGH DO ELEC Electrolyte Solution 125 ml bottle TT TOUGH DO SWM Refi...

Страница 16: ...faulty installations corrosion or the effects of chemical processes is excluded from this warranty coverage Our agents or representatives may provide you with a life expectancy guide based on similar...

Страница 17: ...service and description of the process Description of the fault and circumstances of the failure Process environmental conditions that may be related to the failure of the sensor Statement as to wheth...

Отзывы: