Pressure Compensation Membrane Replacement

Pressure compensation membranes have to be replaced when the sensor has 
been  stored  over  a  longer  period  without  being  kept  in  water  and  the  mem-
branes have lost their elasticity; or when the membranes have been damaged.
When measuring often at great depths, the pressure compensation membrane 
should be checked regularly.

The pressure compensation membrane can be changed as the following describes:

1.  Unscrew the side pressure membrane ring nut with the Pressure Compensator 

Tool TT-TOUGH-DO-SWT or a set of Circlip pliers 

2.  Remove the pressure washer.
3.  Remove the old pressure membrane and the O-Ring.
4.  Refill with electrolyte.

Note:

 If the electrolyte comes into contact with your skin, rinse the affected part 

immediately under running water.

5.  Install the O-Ring and a new pressure compensation membrane.
6.  Refit the pressure washer.
7.  Finally put on the pressure ring nut and screw it on carefully.

Using the Membrane Positioner

Use the Membrane Positioner TT-TOUGH-DO-MP to easily replace the cathode 
membrane without any wrinkles.

Follow these steps for replacing the membrane using a Turtle Tough Membrane 
Positioner:

1.  Pull off the frame.
2.  Put the Positioner on a table.
3.  Centre one membrane on the Positioner.
4.  Press the frame over the Positioner - the membrane gets pre-stretched like 

on a drum.

5.  Hold the Turtle Tough sensor vertically, with the tip upwards.
6.  Put one drop of electrolyte onto the cathode.

Note:

 If the electrolyte comes into contact with your skin, rinse the affected part 

immediately under running water.

7.  Place the Positioner over the sensor - so that the membrane touches the 

cathode - and let it hang down by its own weight.

8.  Gently push down and hold the red collar in its position, then remove the 

Positioner by sliding it down.

9.  Push the red collar down until the groove of the O-Ring is visible (but not 

further!) Continue with replacing the O-Ring and the second red collar as 
usual (see: Replacement of the cathode membrane or Sensor refurbishment).

Содержание TT-TOUGH-DO-M4M

Страница 1: ...TURTLE TOUGH 1 TURTLE TOUGH turtletoughsensors com MANUAL M4M TOUGH DISSOLVED OXYGEN SYSTEM Dissolved Oxygen...

Страница 2: ...emicals 5 Sensors 5 Introduction 6 Specifications 7 Unpacking 7 Operation Temperature Measurement 8 Temperature Stabilisation Time 8 Dissolved Oxygen Measurement 8 Calibration Calibration Verification...

Страница 3: ...eading Safety Precautions the special safety precautions outlined in other sections must also be observed Warnings used in this manual This manual contains vital information relating to the safety of...

Страница 4: ...wired electrical outlet It is recommended that the outlet is protected by an Residual Current Detector or is in any event in compliance with all local electrical regulations Do not bury electrical su...

Страница 5: ...them CAUTION When handling this equipment the accident prevention regulations applicable on site must be observed and the specified personal protective equipment worn PPE examples of protective cloth...

Страница 6: ...d for zero solutions The sensor has an extremely fast response which is critical for applications that demand accuracy or measure at high ppm This is especially true for portable units which are often...

Страница 7: ...atmosphere Housing IP 65 PVC housing Dimensions 200 x 270 x 250 mm H x W x D approx Weight 2 3 kg approx Field case 385 x 355 x 180 mm H x W x D approx Weight 1 5 kg approx Unpacking Please have a cop...

Страница 8: ...antly kept in the measuring liquid it will not be necessary to observe a temperature stabilisation time The best procedure is to wait until the temperature value on the liquid crystal display remains...

Страница 9: ...nd the membrane have to be dry remove any humidity by dabbing the tip with tissue paper Cover the sensor i e with newspaper to avoid any wind on the tip 2 Wait for at least 15 20 minutes to be sure th...

Страница 10: ...material It is held by means of two red collars 4 and one O Ring 5 During measurements the membrane must not come into contact with sharp objects such as stones or sharp grained sand which might dama...

Страница 11: ...verified by holding the sensor horizontally You should now be able to see an air bubble no larger than 4 mm in the centre of the pressure compensation membrane Should refilling of the electrolyte be r...

Страница 12: ...low since the white wick like insert would otherwise be damaged Doing this voids the warranty 6 Wipe the cathode clean with tissue paper but do not polish it because the scratched surface helps to for...

Страница 13: ...ane until electrolyte appears in all the electrolyte channels Note Never use a needle or other sharp objects to create flow since the white wick like insert would otherwise be damaged Doing this voids...

Страница 14: ...pressure ring nut and screw it on carefully Using the Membrane Positioner Use the Membrane Positioner TT TOUGH DO MP to easily replace the cathode membrane without any wrinkles Follow these steps for...

Страница 15: ...Kit for sensor tip with 10 membranes 3 O Rings and 1 bottle electrolyte solution TT TOUGH DO MEM Refill Kit without electrolyte TT TOUGH DO ELEC Electrolyte Solution 125 ml bottle TT TOUGH DO SWM Refi...

Страница 16: ...faulty installations corrosion or the effects of chemical processes is excluded from this warranty coverage Our agents or representatives may provide you with a life expectancy guide based on similar...

Страница 17: ...service and description of the process Description of the fault and circumstances of the failure Process environmental conditions that may be related to the failure of the sensor Statement as to wheth...

Отзывы: