Cyclops Submersible Sensor
998-2100
Rev. 2.9
Page 20
Appendix G: Functional Test for Cyclops with no Connector
To perform a functional check on the Cyclops, connect the Cyclops colored wires to
the power supply and multi-meter as shown in Figure 2 below.
No Connector Cyclops (P/N: 2110/2100-000-
“Identification Letter”-NC)
Additional Equipment required for functional tests:
DC Power Supply, 3 - 15 VDC, >100 mA
Multi-meter to read 0
– 5 VDC
Note: Supply voltages greater than 15 VDC will result in damage to the
sensor.
Figure 2.
With the Cyclops connected as shown in Figure 2 answer questions 1-3 by making
the following functional tests:
1. Is the LED on?
Hold a piece of white paper about ½ an inch in-front of the optical head to
ensure the LED is on.
Note: Cannot perform this test for Turbidity
sensors because they use IR which is not visible.
2. Is there voltage output?
The multi-meter should be reading some voltage >0 VDC
3. Does the voltage output change?
Move the light source closer to your hand or a surface and check if the voltage
output increases
12.00
.015
+
-
DC Power Supply
LED
Light
Output
PSU Positive Connection
(Red)
Supply Ground
0 VDC (Black)
3.52
Multimeter
10 VDC
+
-
Signal Output
“+” (Orange)
Analog Ground
“-” (Green)
Blue
(Tie to Green for 10X)
Yellow
(Tie to Green for 100X)
X1(Low) Gain
Leave both wires disconnected
for Functional Check
Cyclops pins / wires
3
4
2
1
5
6
6
Cyclops pins / wires
3
3
4
4
2
2
1
5
5
6
6
12.00
.015
+
-
DC Power Supply
12.00
.015
+
-
DC Power Supply
LED
Light
Output
LED
Light
Output
PSU Positive Connection
(Red)
Supply Ground
0 VDC (Black)
3.52
Multimeter
10 VDC
+
-
3.52
Multimeter
10 VDC
+
-
Signal Output
“+” (Orange)
Analog Ground
“-” (Green)
Blue
(Tie to Green for 10X)
Yellow
(Tie to Green for 100X)
X1(Low) Gain
Leave both wires disconnected
for Functional Check