Cyclops Submersible Sensor
998-2100
Rev. 2.9
Page 11
4.
Maintenance and Warranty
4.1 Maintenance
4.1.1 Rinsing
The Cyclops should be rinsed or soaked in fresh water following each
deployment, ideally until it is completely clean again.
4.1.2 Care for the bulkhead connector
A light coat of Silicone spray should be used on the rubber of the male
pins of the bulkhead to aid in sealing. The manufacturer recommends
3M™ Silicone Lubricant Spray or Loctite 8021 spray.
Note: You
should avoid using silicone grease. Do NOT use WD-40, it will
destroy the connectors.
4.1.3 Care for the optics
The optical window should be visually inspected after each deployment
following a soaking in fresh water. If cleaning is needed, use optical
tissue to clean the window with soapy water.
Note: The Cyclops should
NOT
come in contact with any organic
solvents (i.e. acetone, methanol) or strong acids and bases.
The UV Cyclops models are the
ONLY
Cyclops sensors that can be
calibrated with Quinine Sulfate standards made in Hydrosulfuric Acid. All
other Cyclops models
CANNOT
be used in
Hydrosulfuric Acid.
4.2 Warranty Terms
Turner Designs warrants the Cyclops and accessories to be free from defects in
materials and workmanship under normal use and service for a period of 12
months from the date of shipment from Turner Designs with the following
restrictions:
Turner Designs is not responsible for replacing parts damaged by
accident or neglect. Your instrument must be installed according to
instructions in the User’s Manual. Damage from corrosion is not
covered. Damage caused by customer modification of the instrument is
not covered.
This warranty covers only Turner Designs products and is not extended
to equipment used with our products. We are not responsible for
incidental or consequential damages, except in those states where this
limitation is not allowed. This warranty gives you specific legal rights and
you may have other rights which vary from state to state.
Damage incurred in shipping is not covered.