2-8
Model 3786 Ultrafine Water-based Condensation Particle Counter
Figure 2-5
Tipping the UWCPC During Drain Cycle for More Complete Drain of the Reservoir
6.
Turn the UWCPC off and remove the power cord.
7.
Remove the Wick.
a.
Loosen the Sample Inlet thumbscrews and remove the
Sample Inlet assembly by pulling downward and out. Do
this near the edge of the table so the growth tube wick can
be removed by sliding it down over the table edge.
b.
Remove the wick from the Inlet assembly. Allow the wick to
air dry and store in a plastic bag with the UWCPC
accessories.
c.
Reinstall the Sample Inlet assembly without the wick and
tighten the thumbscrews.
8.
Drain and dry the reservoir.
a.
Make sure the UWCPC is turned off and the power cord is
removed.
b.
Lay the UWCPC on its side over some paper towel. The Fill
and Drain connectors must be down toward the work
surface.