4-6
Model 8587A Laser Photometer
Table 4-1.
Example Fit Test Sequence
Operation
ASCII CODE
Description of Function
SEND
U
Unlock valves
SEND
P
Set PURGE mode
Delay 20 seconds
Delay for sensor to purge of aerosol
SEND
R
Reset running average
Delay 10 seconds
Average 10 seconds of purge signal data (ZERO)
SEND
D or K
Request data as HEX (D) or Decimal (K)
RECEIVE 8 char
Receive string voltage data
Convert
-
Convert data to decimal long integer (ZERO)
SEND
C
Set UPSTREAM sample mode (chamber)
Delay 20 seconds
Delay for UPSTREAM sample to stabilize
SEND
R
Reset running average
Delay 10 seconds
Average 10 seconds of UPSTREAM (chamber) signal data
SEND
D or K
Request data as HEX (D) or Decimal (K)
RECEIVE 8 char
Receive voltage data
Convert
-
Convert UPSTREAM (chamber) data to decimal long integer
SEND
M
Set DOWNSTREAM sample mode (mask)
SEND
V3F
Set high flow mask purge (mask sample at high flow)
Delay 10 seconds
Delay for mask to purge
SEND
V3N
Set flow for mask sample mode
Delay 20 seconds
Delay for DOWNSTREAM (mask sample) to stabilize
SEND
R
Reset running average
Delay 60 seconds
DOWNSTREAM (mask) signal data averaged over 60 second
interval, since reset.
SEND
D or K
Request averaged data as HEX (D) or Decimal (K)
RECEIVE 8 char
Receive voltage data
Convert
-
Convert DOWNSTREAM (mask) data to decimal long integer
Calculate average
fit factor (see
discussion above)
voltage
ZERO
-
(mask)
voltage
avg.
DOWNSTREAM
voltage
ZERO
-
(chamber)
voltage
UPSTREAM
factor
Fit
F i l t e r T e s t i n g A e r o s o l
Instruments using light scattering such as the Model 8587A are widely
specified for filter testing, particularly since good correlation exists between
light scattering and a mass measurement. TSI uses the same aerosol
sensor in the Model 8587A Laser Photometer as is used in its popular
Model 8130 Filter Tester.
Aerosols of Dioctyl-Phalate (DOP), Emery Oil, or Olive oil or a mean
diameter of 0.2 to 0.3 microns and geometric standard deviation (GSD) of
approximately 1.6 or less are typically used for filter testing.
Particle size is very important for filter testing since test results will depend
on the size used. The aerosol size indicated above is desirable in that it is
in the most penetrating particle size range for typical non-electret filters.
Содержание 8587A
Страница 1: ...LASER PHOTOMETER MODEL 8587A OPERATION AND SERVICE MANUAL P N 1980538 REVISION D JULY 2015 ...
Страница 2: ......
Страница 10: ...x Model 8587A Laser Photometer This page intentionally left blank ...
Страница 16: ...xvi Model 8587A Laser Photometer This page intentionally left blank ...
Страница 22: ...2 4 Model 8587A Laser Photometer This page intentionally left blank ...
Страница 34: ...3 12 Model 8587A Laser Photometer Figure 3 4 Sheath Air Flow System ...
Страница 46: ...4 8 Model 8587A Laser Photometer This page intentionally left blank ...
Страница 52: ...5 6 Model 8587A Laser Photometer This page intentionally left blank ...
Страница 68: ...B 14 Model 8587A Laser Photometer This page intentionally left blank ...
Страница 80: ...C 12 Model 8587A Laser Photometer This page intentionally left blank ...
Страница 84: ......