3-10
Model 8587A Laser Photometer
N o t e
Stable airflows are extremely important in obtaining repeatable test
results. TSI strongly suggests the instrument be operated with a vacuum
greater than 55 kPa (8.0 psi). At a vacuum of greater than 55 kPa
(8.0 psi) an internal critical orifice limits and stabilizes the sample airflow
to 2.0 L/min (0.07 cfm). Repeatable test results will ONLY be obtained
with stable sample airflows.
For the PURGE mode, the pump must provide a
free flow
of between 12 to
30 L/min (0.43 to 1.1 cfm).
The vacuum pump is connected to the port labeled VACUUM on the back
panel. Refer to the
Chapter 4, “
” for more information
on pump requirements.
Figure 3-3
Flow Schematic Model 8587A Laser Photometer