Press the up or down arrow keys to toggle between 1 and 2 stop bits. When the display
shows the desired choice, press "FUNC/ENTR" to enter all the serial port data.
3.20
FUNCTION 05 - TIME QUALITY ENABLE/SETUP
3.20.1
Both the front-panel numeric display and the Serial I/O time output string indicate time
quality. Refer to Serial I/O FUNCTION FO8 - CONTINUOUS TIME ONCE PER
SECOND for a description of the time quality indication in the Serial I/O time output
string. As shipped the time quality indication is enabled and the thresholds are:
First threshold
10
µ
s
Second threshold
100
µ
s
Third threshold
1000
µ
s
Fourth threshold
10000
µ
s
3.20.2
Use Function 05 to enable or disable the time quality indication or set the worst-case-
error thresholds. Press "FUNC/ENTR", then "0" "5". The display will show:
Time quality
or
Time quality
on
off
3.20.3
Press the up or down arrow keys to toggle between "on" and "off". When the display
shows the desired state, press "FUNC/ENTR". The display will show:
First tq flag:
00000010000ns
3.20.4
or the current value. Press the right or left arrow keys to position the cursor beneath the
digit that you wish to change. Press the up or down arrow keys to scroll through the
possible digit choices. Alternately, directly enter the numbers using the key pad. The
cursor will advance to the next digit automatically. When the display shows the value
that you desire for the first time quality threshold, press "FUNC/ENTR". The display will
show:
Second tq flag:
00000100000ns
or the current value. Press the right or left arrow keys to position the cursor beneath the
digit that you wish to change. Press the up or down arrow keys to scroll through the
possible digit choices. Alternately, directly enter the numbers using the key pad. The
cursor will advance to the next digit automatically. When the display shows the value that
you desire for the second time quality threshold, press "FUNC/ENTR". The display will
show:
Third tq flag:
00001000000ns
or the current value. Press the right or left arrow keys to position the cursor beneath the
digit that you wish to change. Press the up or down arrow keys to scroll through the
possible digit choices. Alternately, directly enter the numbers using the key pad. The
cursor will advance to the next digit automatically. When the display shows the value that
your desire for the third time quality threshold, press "FUNC/ENTR". The display will show: